COBAS INTEGRA 400
Report
- Report Number
- 1823260-2008-07107
- Event Type
- Malfunction
- Date Received
- September 23, 2008
- Date of Event
- September 4, 2008
- Report Date
- September 23, 2008
- Manufacturer
- ROCHE DIAGNOSTICS CORP.
- Product Code
- JJE
- PMA / PMN Number
- K951595
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 133 MMOL/L, REPEAT 140 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.
USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 127 MMOL/L, REPEAT 134 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.
USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 131 MMOL/L, REPEAT 139 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.
USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 133 MMOL/L, REPEAT 142 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS INTEGRA 400 | CLINICAL CHEMISTRY ANALYZER - JJE | JJE | ROCHE DIAGNOSTICS CORP. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | |||
| 2 | 84 YR | |||
| 3 | UNK | |||
| 4 | 69 YR |