FDA Adverse Event Malfunction Summary report: N

COBAS INTEGRA 400

MDR report key: 1180817 · Received September 23, 2008

Report

Report Number
1823260-2008-07107
Event Type
Malfunction
Date Received
September 23, 2008
Date of Event
September 4, 2008
Report Date
September 23, 2008
Manufacturer
ROCHE DIAGNOSTICS CORP.
Product Code
JJE
PMA / PMN Number
K951595
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Description of Event or Problem · 1

USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 133 MMOL/L, REPEAT 140 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.

Description of Event or Problem · 2

USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 127 MMOL/L, REPEAT 134 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.

Description of Event or Problem · 3

USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 131 MMOL/L, REPEAT 139 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.

Description of Event or Problem · 4

USER RECEIVING DISCREPANT RESULTS FOR SODIUM. FOUR PATIENT SAMPLES ARE AFFECTED. REPEAT TESTING WAS PERFORMED ON OTHER ANALYZER. INITIAL RESULT 133 MMOL/L, REPEAT 142 MMOL/L. REPEAT RESULTS WERE REPORTED FOR ALL SAMPLES. THE FIELD SERVICE REP DETERMINED MIX DRY PRESSURE WAS LOW AND THE MIX TOWER WAS CLOGGED. HE REPLACED THE MIX TOWER AND ADJUSTED THE MIX DRY PRESSURE. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS INTEGRA 400 CLINICAL CHEMISTRY ANALYZER - JJE JJE ROCHE DIAGNOSTICS CORP.

Patients

Seq Age Sex Outcome Treatment
1 79 YR
2 84 YR
3 UNK
4 69 YR