FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN MANAGEMENT SYSTEM

MDR report key: 11807010 · Received May 11, 2021

Report

Report Number
3004464228-2021-06987
Event Type
Malfunction
Date Received
May 11, 2021
Date of Event
April 27, 2021
Report Date
April 27, 2021
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
20385081120033
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED/RECEIVED TO DATE. IF THE DEVICE IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE INVESTIGATION RESULTS. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING CONFIRMING THE BLUE SOFT CANNULA IS INSPECTED FOR ANY DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVELS REACHED OVER 300 MG/DL WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS. WHEN REMOVED FROM THE INFUSION SITE (ARM), THE POD'S CANNULA WAS FOUND BENT. LIGHT BLEEDING AT THE SITE WAS ALSO REPORTED. AS TREATMENT, A NEW POD WOULD SOON BE APPLIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
704850 OMNIPOD INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 19191 L72182 20385081120033

Patients

Seq Age Sex Outcome Treatment
1 28 YR