OMNIPOD INSULIN MANAGEMENT SYSTEM
Report
- Report Number
- 3004464228-2021-06987
- Event Type
- Malfunction
- Date Received
- May 11, 2021
- Date of Event
- April 27, 2021
- Report Date
- April 27, 2021
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- UDI-DI
- 20385081120033
- PMA / PMN Number
- K192659
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS NOT BEEN RETURNED/RECEIVED TO DATE. IF THE DEVICE IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE INVESTIGATION RESULTS. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING CONFIRMING THE BLUE SOFT CANNULA IS INSPECTED FOR ANY DAMAGE.
IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVELS REACHED OVER 300 MG/DL WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS. WHEN REMOVED FROM THE INFUSION SITE (ARM), THE POD'S CANNULA WAS FOUND BENT. LIGHT BLEEDING AT THE SITE WAS ALSO REPORTED. AS TREATMENT, A NEW POD WOULD SOON BE APPLIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 704850 | OMNIPOD INSULIN MANAGEMENT SYSTEM | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 19191 | L72182 | 20385081120033 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |