FDA Adverse Event Malfunction Summary report: N

TRUE METRIX AIR

MDR report key: 11806338 · Received May 11, 2021

Report

Report Number
1000113657-2021-00310
Event Type
Malfunction
Date Received
May 11, 2021
Date of Event
April 15, 2021
Report Date
May 11, 2021
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). TEST STRIPS WERE NOT RETURNED FOR EVALUATION. METER WAS RETURNED FOR EVALUATION - REPORTED DEFECT NOT REPRODUCED. NO DEFECT FOUND MOST LIKELY UNDERLYING ROOT CAUSE MLC-058 USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 21-APR-2021 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM BACK TO BACK BLOOD TESTS OF 133, 211, 214 AND 227MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 120MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 228MG/DL AND 222MG/DL USING TRUE METRIX AIR METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 06/21/2022 AND TEST STRIPS WERE OPENED FOUR DAYS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY:(B)(6).