FDA Adverse Event Injury Summary report: N

STARCLOSE VASCULAR CLOSURE SYSTEM

MDR report key: 1180461 · Received September 26, 2008

Report

Report Number
2953144-2008-01573
Event Type
Injury
Date Received
September 26, 2008
Date of Event
September 1, 2008
Report Date
September 11, 2008
Manufacturer
ABBOTT VASCULAR REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR EVAL. IT HAS NOT YET BEEN RECEIVED.

Description of Event or Problem · 1

DEVICE MALFUNCTION: DIFFICULT TO REMOVE. TIME OF DEVICE MALFUNCTION: DURING VESSEL CLOSURE. SYMPTOMS/AE: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF A FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, THE CLIP WAS DEPLOYED; HOWEVER, THE DEVICE WAS DIFFICULT TO REMOVE FROM THE PATIENT'S ARTERY. THE DEVICE WAS REMOVED USING COUNTER-TRACTION AND AN ASSERTIVE PULL PER THE INSTRUCTION FOR USE (IFU). HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WERE NO REPORTED ADVERSE PT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFO WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR REDWOOD CITY NA 66079-6H

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention