FDA Adverse Event Other Summary report: N

BECTON DICKINSON 10 ML SYRINGE LUER-LOK TIP

MDR report key: 1180447 · Received September 29, 2008

Report

Report Number
MW5008474
Event Type
Other
Date Received
September 29, 2008
Date of Event
September 25, 2008
Report Date
September 29, 2008
Manufacturer
BECTON DICKINSON
Product Code
FMF
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
PHARMACIST

Narratives

Description of Event or Problem · 1

FLUID FOUND ON THE SYRINGE PUMP. SYRINGE OF DOPAMINE CHECKED AND FOUND TO HAVE SEVERAL ML OF AIR AND ONLY 2 ML OF DOPAMINE. THE DOPAMINE HAD BEEN STARTED AT 1600 AND HAD 8 ML OF DOPAMINE WHEN THE INFUSION WAS STARTED AT A RATE LESS THAN 0.5 ML/HR. UPON DISCOVERY THE NURSE CLAMPED THE LINE. AIR HAD INFUSED ONE THIRD OF THE LENGTH OF THE IV LINE MOVING TOWARDS THE PT. BLOOD PRESSURE HAD BEEN LOW SINCE THE START OF THE INFUSION WITH SEVERAL CHANGES IN DOSAGE POSSIBLY DUE TO THE SYRINGE LEAK. SYRINGE AND PUMP REMOVED FROM BEDSIDE. SYRINGE PUMP SENT TO BIOMEDICAL AND SYRINGE FOUND TO HAVE A LARGE CRACK ALONG THE SIDE. THE INPATIENT PHARMACY WAS NOTIFIED AS WAS THE BECTON DICKINSON REPRESENTATIVE. THE PHARMACY IDENTIFIED AND SEQUESTERED THE LOT NUMBER FROM WHERE THE SYRINGE WAS LIKELY TO HAVE COME. MOST OF THE SYRINGES WERE GIVEN TO THE BD REPRESENTATIVE FOR TESTING. PHARMACY, NURSING AND MATERIALS STAFF NOTIFIED OF THE CONCERN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECTON DICKINSON 10 ML SYRINGE LUER-LOK TIP 10 ML SYRINGE FMF BECTON DICKINSON REF 309604 8151405

Patients

Seq Age Sex Outcome Treatment
1 18 YR MEDFUSION PUMP 3500