CARD SYNERGY PLUS/PULSE
Report
- Report Number
- 1045510-2008-00017
- Event Type
- Other
- Date Received
- September 29, 2008
- Date of Event
- November 15, 2007
- Report Date
- September 26, 2008
- Manufacturer
- HILL-ROM CHARLESTON
- Product Code
- IKY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
RESULTS OF THE HILL-ROM ENGINEERING INVESTIGATION ARE AS FOLLOWS: OUTPUT OF AIR REACHED 146.8 DEGREES FAHRENHEIT AFTER 1.5 HOURS OF OPERATION. THE AIR MANIFOLD WAS TAPED IN PLACE, THE PLATE TO CONTROL THE BLOWER MOTOR COOLING AIR WAS NOT SECURE AND THE HEATED AIR OFF OF THE MOTOR WAS DISCHARGED WITHIN THE BLOWER COMPARTMENT AND BEING PULLED INTO THE INTAKE OF THE BLOWER, THE O RINGS OF THE MANIFOLD PLUG WERE COMPRESSED AND TOOK A SET, THIS ALLOWED THE PLUG NOT TO BE SECURE AND TIGHT WHICH ALLOWS HEATED AIR TO RECIRCULATE IN THE BLOWER COMPARTMENT, FITTINGS AND TUBING ON THE MANIFOLD WERE NOT SECURE AND TIGHT WHICH LEAKED AIR, A HOSE FITTING WAS DEFORMED DUE TO HEAT WHICH ALLOWED IT TO LEAK AIR. ALL OF THESE CONDITIONS COULD HAVE CONTRIBUTED TO THE EXTERNAL AIR OUTPUT REACHING A TEMPERATURE OF 146.9 DEGREES FAHRENHEIT.
IT WAS REPORTED TO HILL-ROM THAT THE MATTRESS WAS BURNING THE PATIENT'S FOOT. A HILL-ROM TECHNICIAN FOUND THAT THE UNIT WAS WORKING WITHIN SPECIFICATION. A FURTHER INVESTIGATION OF THE UNIT FOUND THAT THE OUTPUT OF THE AIR REACHED 146.8 DEGREES FAHRENHEIT AFTER 1.5 HOURS OF OPERATION. DURING AN INVESTIGATION OF THIS PRODUCT LINE IT WAS FOUND THAT IS IT POSSIBLE FOR AN EXTERNAL COMPONENT ON THE PRODUCT TO REACH A TEMPERATURE ABOVE THAT WHICH IS ALLOWABLE FOR BRIEF PATIENT CONTACT. ALL ALLEGATIONS OF INJURY (REGARDLESS OF SEVERITY) DUE TO THIS FAILURE MODE ARE BEING REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARD SYNERGY PLUS/PULSE | SYNERGY PULSE | IKY | HILL-ROM CHARLESTON | 5609 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |