RIGHT ANGLE LINEAR CUTTER, 45MM DLU
MDR report key: 1180214
·
Received September 29, 2008
Report
- Report Number
- 2532140-2008-00067
- Event Type
- Injury
- Date Received
- September 29, 2008
- Date of Event
- September 5, 2008
- Report Date
- September 5, 2008
- Manufacturer
- POWER MEDICAL INTERVENTIONS
- Product Code
- GDW
- PMA / PMN Number
- K012809
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
AFTER ANALYSIS, IT APPEARS THAT THE FIRE SIDE DRIVE CLIP OF THE RALC45 WAS BENT INWARDS. THIS MAY HAVE OCCURRED WHEN THE RALC45 WAS ATTACHED TO THE IDRIVE.
Description of Event or Problem · 1
A RALC45 DEVICE WAS CONNECTED TO AN IDRIVEC DEVICE. THE RALC DEVICE WAS CALIBRATED, POSITIONED, THEN CLOSED AND FIRED. A GRINDING NOISE WAS HEARD, AFTER FIRING SEQUENCE, THE SURGEON NOTED THAT THE TISSUE WAS PARTIALLY CUT AND THAT NO STAPLES WERE PLACED IN THE TISSUE. THE LEAK WAS CLOSED WITH SUTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIGHT ANGLE LINEAR CUTTER, 45MM DLU | STAPLE, IMPLANTABLE | GDW | POWER MEDICAL INTERVENTIONS | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |