FDA Adverse Event
Injury
Summary report: N
INTELECT ADVANCED STIM MONOCHROME
MDR report key: 1180211
·
Received September 29, 2008
Report
- Report Number
- 1022819-2008-00226
- Event Type
- Injury
- Date Received
- September 29, 2008
- Date of Event
- March 20, 2007
- Report Date
- March 20, 2007
- Manufacturer
- CHATTANOOGA GROUP
- Product Code
- IPF
- Removal / Correction Number
- 1022819-11/15/2007-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THIS DEVICE IS FOR EXPORT ONLY. DEVICE 2: BRAND NAME: VACUUM MODULE W/CART PKG. DEVICE NAME: VACUUM ELECTRODE MODULE. MODEL #: 2774, CATALOG #: 2774, SERIAL #: (B)(4).
Description of Event or Problem · 1
UNIT SHOCKED STAFF MEMBER OF HEALTH FACILITY AFTER INSTALLATION OF THE DEVICE. RESULTED IN BURN. THIS MALFUNCTION CAUSES THE PRINTED CIRCUIT BOARD TO CHANGE THE STIMULATORY PT OUTPUT FROM THE INTENDED OUTPUT TO AN UNINTENDED HIGHER VOLTAGE OUTPUT. WHEN SUCH A MALFUNCTION OCCURS, IT RESULTS IN A SHOCK SENSATION AND POSSIBLE BURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTELECT ADVANCED STIM MONOCHROME | IPF, GZJ, GZI | IPF | CHATTANOOGA GROUP | 2773MS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |