FDA Adverse Event
Malfunction
Summary report: N
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 11797808
·
Received May 10, 2021
Report
- Report Number
- 3004753838-2021-83067
- Event Type
- Malfunction
- Date Received
- May 10, 2021
- Date of Event
- April 12, 2021
- Report Date
- May 10, 2021
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
COM-0953573-21
Description of Event or Problem · 1
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. IT WAS DETERMINED THAT THE SIGNAL LOSS WAS RELATED TO THE MOBILE APPLICATION. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS OFF-LABEL USAGE OF THE DEVICE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE PATIENT CLOSED THE MOBILE APP. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692259 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | DEXCOM, INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |