FDA Adverse Event Malfunction Summary report: N

VIDEO SYSTEM CENTER

MDR report key: 11795910 · Received May 10, 2021

Report

Report Number
8010047-2021-06062
Event Type
Malfunction
Date Received
May 10, 2021
Date of Event
April 14, 2021
Report Date
June 30, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FET
PMA / PMN Number
K122831
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC), BUT RETURNED TO OLYMPUS REPAIR CENTER FOR EVALUATION. ACCORDING TO THE EVALUATION, THE FOLLOWING WAS FOUND. DVI OUTPUT DID NOT WORK INTERMITTENTLY DUE TO A FAILURE OF THE INNER CIRCUIT BOARD. THERE WERE MINOR SCRATCHES ON THE TOP COVER, CHASSIS, AND FRONT PANEL. THE SOCKET WAS RUSTED AND PEELED OFF. DEVICE HISTORY RECORD REVIEW INDICATES THAT THE PRODUCT WAS MANUFACTURED AND TESTED IN ACCORDANCE WITH ALL APPLICABLE PROCEDURES AND MET ALL FINAL PRODUCT RELEASE CRITERIA. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. OMSC SURMISED THAT THIS PHENOMENON ATTRIBUTED TO A FAILURE OF THE INNER CIRCUIT BOARD. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THE DEVICE IN THIS REPORT HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

DURING THE PREPARATION FOR USE, THE USER FOUND THE FOLLOWING. THE FRONT PANEL KEYBOARD DID NOT WORK. THE LAMP DID NOT LIGHT. ALL LEDS OF THE FRONT PANEL WERE LIGHTED. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT. THE USER FACILITY DID NOT PROVIDE OTHER DETAILED INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693493 VIDEO SYSTEM CENTER VIDEO SYSTEM CENTER FET OLYMPUS MEDICAL SYSTEMS CORP. CV-170

Patients

Seq Age Sex Outcome Treatment
1