VIDEO SYSTEM CENTER
Report
- Report Number
- 8010047-2021-06062
- Event Type
- Malfunction
- Date Received
- May 10, 2021
- Date of Event
- April 14, 2021
- Report Date
- June 30, 2021
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FET
- PMA / PMN Number
- K122831
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE DEVICE WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC), BUT RETURNED TO OLYMPUS REPAIR CENTER FOR EVALUATION. ACCORDING TO THE EVALUATION, THE FOLLOWING WAS FOUND. DVI OUTPUT DID NOT WORK INTERMITTENTLY DUE TO A FAILURE OF THE INNER CIRCUIT BOARD. THERE WERE MINOR SCRATCHES ON THE TOP COVER, CHASSIS, AND FRONT PANEL. THE SOCKET WAS RUSTED AND PEELED OFF. DEVICE HISTORY RECORD REVIEW INDICATES THAT THE PRODUCT WAS MANUFACTURED AND TESTED IN ACCORDANCE WITH ALL APPLICABLE PROCEDURES AND MET ALL FINAL PRODUCT RELEASE CRITERIA. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. OMSC SURMISED THAT THIS PHENOMENON ATTRIBUTED TO A FAILURE OF THE INNER CIRCUIT BOARD. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
THE DEVICE IN THIS REPORT HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.
DURING THE PREPARATION FOR USE, THE USER FOUND THE FOLLOWING. THE FRONT PANEL KEYBOARD DID NOT WORK. THE LAMP DID NOT LIGHT. ALL LEDS OF THE FRONT PANEL WERE LIGHTED. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT. THE USER FACILITY DID NOT PROVIDE OTHER DETAILED INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 693493 | VIDEO SYSTEM CENTER | VIDEO SYSTEM CENTER | FET | OLYMPUS MEDICAL SYSTEMS CORP. | CV-170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |