ID NOW COVID-19
Report
- Report Number
- 1221359-2021-01445
- Event Type
- Malfunction
- Date Received
- May 7, 2021
- Date of Event
- April 1, 2021
- Report Date
- September 19, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCABOROUGH
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION NOT YET COMPLETE. UPON COMPLETION, THE INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 1023941 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1023941 AND TEST BASE PART NUMBER 190-430 / LOT 1023941. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1023941 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY.
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED NASAL KITTED SWAB. THE PATIENT GOT TESTED ELSEWHERE AND TESTED NEGATIVE. THE CUSTOMER STATED THE PATIENT WAS SYMPTOMATIC WITH FATIGUE. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 689802 | ID NOW COVID-19 | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCABOROUGH | 1023941 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |