FDA Adverse Event Malfunction Summary report: N

SMR - EXPANSION EXTRACTOR

MDR report key: 11789067 · Received May 7, 2021

Report

Report Number
3008021110-2021-00035
Event Type
Malfunction
Date Received
May 7, 2021
Date of Event
April 28, 2021
Report Date
September 16, 2024
Manufacturer
LIMACORPORATE S.P.A.
Product Code
KWT
PMA / PMN Number
K100858
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: BY CHECKING THE DHR OF THE LOT NUMBER: 18AQ1G4, NO PRE-EXISTING ANOMALIES WERE DETECTED ON THE PIECES MANUFACTURED WITH THIS LOT NUMBER. THIS IS THE FIRST AND ONLY COMPLAINT RECEIVED ON THIS LOT NUMBER. THE INSTRUMENT WAS NOT AVAILABLE TO BE RETURNED FOR INSPECTION. THEREFORE, BASED ON THE VERY FEW INFORMATION RECEIVED, WE ARE NOT ABLE TO FURTHER INVESTIGATE THE EVENT. HOWEVER, CONSIDERING THAT: CHECKING THE MANUFACTURING CHARTS OF THE LOT NUMBER INVOLVED IN THE EVENT, NO PRE-EXISTING ANOMALY WAS FOUND OUT. WE HAVE NO EVIDENCE TO CONSIDER THE EVENT AS PRODUCT RELATED. PMS DATA: ACCORDING TO THE RELEVANT PMS DATA, WE CAN ESTIMATE THE OCCURRENCE RATE OF MALFUNCTIONING OF THE SMR - EXPANSION ESTRACTOR (PART CODE 9013.52.165) TO BE AROUND (B)(4) WORLDWIDE. PLEASE NOTE THAT THIS OCCURRENCE RATE IS OVERESTIMATED BECAUSE IT DOES NOT CONSIDER THE REUSE OF THE INSTRUMENTS BUT ONLY THE TOTAL NUMBER OF PIECES MANUFACTURED. BASED ON THE ROOT CAUSE ANALYSIS PERFORMED AND ACCORDING TO THE RELEVANT PMS DATA, NO CORRECTIVE ACTIONS REQUIRED FOR THIS SPECIFIC CASE. LIMACORPORATE WILL CONTINUE MONITORING THE MARKET TO PROMPTLY DETECT ANY FURTHER SIMILAR ISSUE. NOTE: THIS IS A FINAL MDR.

Description of Event or Problem · 0

DURING SHOULDER SURGERY PERFORMED ON (B)(6) 2021, THE STEM FROM THE PROXIMAL HUMERAL BODY WAS GOING TO BE EXTRACTED BUT THE SMR - EXPANSION EXTRACTOR (PART CODE: 9013.52.165, LOT NUMBER: 18AQ1G4) WOULD NOT FIT FAR ENOUGH ON TO EXTRACT OFF. SURGERY WAS PROLONGED OF 20 MINUTES. EVENT HAPPENED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

BY CHECKING THE DHR OF THE LOT #18AQ1G4, NO PRE-EXISTING ANOMALIES WERE DETECTED ON THE PIECES MANUFACTURED WITH THIS LOT #. THIS IS THE FIRST AND ONLY COMPLAINT RECEIVED ON THIS LOT#. WE WILL SUBMIT A FINAL MDR ONCE THE INVESTIGATION WILL BE COMPLETED.

Description of Event or Problem · 1

DURING SHOULDER SURGERY PERFORMED ON (B)(6) 2021, THE STEM FROM THE PROXIMAL HUMERAL BODY WAS GOING TO BE EXTRACTED BUT THE SMR - EXPANSION EXTRACTOR (PRODUCT CODE 9013.52.165, LOT# 18AQ1G4) WOULD NOT FIT FAR ENOUGH ON TO EXTRACT OFF. SURGERY WAS PROLONGED OF 20 MINUTES. EVENT HAPPENED IN THE US.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
691248 SMR - EXPANSION EXTRACTOR EXPANSION EXTRACTOR KWT LIMACORPORATE S.P.A. 9013.52.165 18AQ1G4

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other