ID NOW COVID-19
Report
- Report Number
- 1221359-2021-01424
- Event Type
- Malfunction
- Date Received
- May 6, 2021
- Date of Event
- April 1, 2021
- Report Date
- September 19, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCABOROUGH
- Product Code
- QJR
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION NOT YET COMPLETE. UPON COMPLETION, THE INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. REFERENCE MFR. REPORTS: 1221359-2021-01423, 1221359-2021-01425, 1221359-2021-01426.
INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 101747 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1017647, TEST BASE PART NUMBER 190-430 / LOT 1017647. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 101747 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REVIEW OF COMPLAINTS AGAINST THE KIT LOTS FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY.
THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS PERFORMED ON VARIOUS. THIS MFR. REPORT ADDRESSES PATIENT ONE (2) OF FOUR (4). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED NASAL SWAB ( PURITAN STERILE FOAM TIPPED APPLICATOR). REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING (X2) ON NASOPHARYNGEAL SWABS WITH PCR GENERATED NEGATIVE RESULTS. CT VALUES NOT PROVIDED. THE PATIENT WAS NOT SYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. THE CUSTOMER REPORTED THE PATIENT MISSED TWO DAYS OF WORK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 684444 | ID NOW COVID-19 | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCABOROUGH | 101747 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |