FDA Adverse Event Malfunction Summary report: N

ULTHERA SYSTEM

MDR report key: 11787534 · Received May 6, 2021

Report

Report Number
3013840437-2021-00120
Event Type
Malfunction
Date Received
May 6, 2021
Date of Event
April 8, 2021
Report Date
April 8, 2021
Manufacturer
MERZ NORTH AMERICA, INC.
Product Code
OHV
UDI-DI
00840763100179
PMA / PMN Number
K180623
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE ULTHERA CONTROL UNIT AND ASSOCIATED POWER CORD WERE RETURNED TO MERZ ON 23-APR-2021. A PRELIMINARY INVESTIGATION WAS PERFORMED BY A MERZ SENIOR BIOMEDICAL ENGINEER ON 29-APR-2021. DURING THIS INVESTIGATION, NO DAMAGE, EVIDENCE OF CHARRING, OR OTHER DEVIATIONS WERE OBSERVED WITH THE CONTROL UNIT NOR THE POWER CORD. THE UNIT WAS EVALUATED AND WAS FOUND TO BE IN NORMAL WORKING CONDITION. IN AN ATTEMPT TO REPLICATE THE ISSUE, THE POWER CORD WAS PARTIALLY INSERTED INTO THE CONTROL UNIT AND MINOR SPARKING WAS OBSERVED. THIS CONDITION MAY BE OBSERVED ON MOST AC POWERED DEVICES WHEN THERE IS A POOR CONNECTION BETWEEN A POWER SUPPLY AND THE DEVICE. THIS CONTROL UNIT HAS BEEN SENT FOR COMPLETE DIAGNOSTIC EVALUATION AND TESTING. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. WHEN THE SERVICE INVESTIGATION HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING A PHONE CALL WITH A PRACTICE FOR AN UNRELATED EVENT ON (B)(6) 2021, MERZ ULTHERA RECEIVED INFORMATION AN ULTHERA CONTROL UNIT WAS EMITTING SPARKS FROM THE BACK OF THE UNIT WHERE THE POWER CORD PLUGS INTO THE BACK OF THE UNIT. THIS ISSUE OCCURRED WHEN THE USER MOVED THIS DEVICE TO ANOTHER LOCATION WITHIN THE PRACTICE AND PLUGGED IT INTO THE WALL OUTLET. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680192 ULTHERA SYSTEM ULTHERA SYSTEM OHV MERZ NORTH AMERICA, INC. UC-1 00840763100179

Patients

Seq Age Sex Outcome Treatment
1 Other