FDA Adverse Event Injury Summary report: N

RETROGRADE DRL 10MM

MDR report key: 11786794 · Received May 6, 2021

Report

Report Number
1219602-2021-01045
Event Type
Injury
Date Received
May 6, 2021
Date of Event
March 16, 2021
Report Date
August 13, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GFG
UDI-DI
00885554037753
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H10: H3, H6: THE REPORTED DEVICE, USED IN TREATMENT, WAS NOT RETURNED TO THE DESIGNATED COMPLAINT UNIT FOR INDEPENDENT EVALUATION, THUS VISUAL INSPECTION AND FUNCTIONAL TESTING COULD NOT BE PERFORMED. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW CONCLUDED THIS WAS AN ISOLATED ISSUE. THE INSTRUCTIONS FOR USE WAS REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. A REVIEW OF RISK MANAGEMENT FILES FOUND THAT THE REPORTED FAILURE WAS DOCUMENTED APPROPRIATELY. A RELATIONSHIP, IF ANY, BETWEEN THE SUBJECT DEVICE AND THE REPORTED EVENT COULD NOT BE DETERMINED. PLEASE REFER TO THE INSTRUCTIONS FOR USE FOR RECOMMENDATIONS ON PROPER USE OF THE DEVICE AND POTENTIAL TROUBLESHOOTING METHODS TO PREVENT FUTURE REOCCURRENCE OF THE REPORTED EVENT. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME. IF THE PRODUCT ASSOCIATED WITH THIS EVENT IS RETURNED AT A FUTURE DATE, THIS INVESTIGATION WILL BE REOPENED FOR EVALUATION.

Additional Manufacturer Narrative · 1

INTERNAL COMPLAINT REFERENCE CASE-(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING THE PROCEDURE, THE "10MM TRUNAV RETROGRADE DRILL BIT" BROKE INTO 2 PIECES. IT IS UNKNOWN IF A BACKUP DEVICE WAS AVAILABLE OR IF THERE WAS A DELAY IN THE SURGICAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681967 RETROGRADE DRL 10MM BIT, SURGICAL GFG SMITH & NEPHEW, INC. 72204046 4905288 00885554037753

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention