FDA Adverse Event Injury Summary report: N

LIGHT ADJUSTABLE LENS (LAL0

MDR report key: 11784987 · Received May 6, 2021

Report

Report Number
3012712027-2021-00020
Event Type
Injury
Date Received
May 6, 2021
Date of Event
April 6, 2021
Report Date
May 6, 2021
Manufacturer
RXSIGHT, INC.
Product Code
PZK
UDI-DI
00818806020159
PMA / PMN Number
P160055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

SITE REPORTED THAT PATIENT WAS INITIALLY IMPLANTED WITH THE LIGHT ADJUSTABLE LENS AND HAD A GOOD OUTCOME. THE PATIENT THEN REPORTED CLOUDY VISION. RETROILLUMINATION OF THE EYE VERIFIED PREMATURE PHOTOPOLYMERIZATION OF THE LENS. THE LENS WAS EXPLANTED ON (B)(6) 2021. THE EXPLANTED LENS WAS RETURNED FOR EVALUATION. INSPECTION OF THE RETURNED LENS CONFIRMED THE PRESENCE OF A ZONE IN THE CENTER OF THE LENS, WHICH IS INDICATIVE OF PREMATURE PHOTOPOLYMERIZATION OF THE LENS.

Description of Event or Problem · 0

SITE REPORTED THAT PATIENT WAS INITIALLY IMPLANTED WITH THE LIGHT ADJUSTABLE LENS AND HAD A GOOD OUTCOME. THE PATIENT THEN REPORTED CLOUDY VISION. RETROILLUMINATION OF THE EYE VERIFIED PREMATURE PHOTOPOLYMERIZATION OF THE LENS. THE LENS WAS EXPLANTED ON (B)(6) 2021.

Additional Manufacturer Narrative · 1

SITE REPORTED THAT PATIENT WAS INITIALLY IMPLANTED WITH THE LIGHT ADJUSTABLE LENS AND HAD A GOOD OUTCOME. THE PATIENT THEN REPORTED CLOUDY VISION. RETROILLUMINATION OF THE EYE VERIFIED PREMATURE PHOTOPOLYMERIZATION OF THE LENS. THE LENS WAS EXPLANTED ON (B)(6) 2021. DEVICE HISTORY RECORD FOR THE LENS WAS REVIEWED. NO ISSUES WERE NOTED. THE EXPLANTED LENS HAS BEEN RETURNED AND IS CURRENTLY BEING EVALUATED.

Description of Event or Problem · 1

SITE REPORTED THAT PATIENT WAS INITIALLY IMPLANTED WITH THE LIGHT ADJUSTABLE LENS AND HAD A GOOD OUTCOME. THE PATIENT THEN REPORTED CLOUDY VISION. RETROILLUMINATION OF THE EYE VERIFIED PREMATURE PHOTOPOLYMERIZATION OF THE LENS. THE LENS WAS EXPLANTED ON (B)(6) 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680500 LIGHT ADJUSTABLE LENS (LAL0 LIGHT ADJUSTABLE LENS (LAL) PZK RXSIGHT, INC. 60005 L02-001349 00818806020159

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention