FDA Adverse Event Malfunction Summary report: N

C500 CORPUS

MDR report key: 11783282 · Received May 6, 2021

Report

Report Number
1221084-2021-00018
Event Type
Malfunction
Date Received
May 6, 2021
Date of Event
December 28, 2020
Report Date
May 6, 2021
Manufacturer
PERMOBIL AB (PAB)
Product Code
ITI
PMA / PMN Number
K991658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

EVALUATION OF THE DEVICE CONFIRMED THE TOP PLATE THAT SECURES THE SEATING TO THE ELEVATOR HAVING DETACHED FROM WHERE IT ATTACHED TO THE ELEVATOR INNER TUBE. WHEN THIS OCCURED, IT REPORTEDLY CAUSED THE SEATING TO FALL OVER TO THE SIDE LANDING ON THE GROUND. REPORTS INDICATE THE END-USER DID NOT INCUR ANY SERIOUS INJURIES, HAVING SUFFERED MINOR BRUISING TO THEIR SHOULDER THAT DID NOT NECESSITATE MEDICAL INTERVENTION. THE COMPONENT FAILURE IS BEING ATTRIBUTED TO STRESSES BEING APPLIED OVER TIME LEADING TO EVENTUAL MATERIAL FATIGUE OF THE TOP PLATE. A CORRECTIVE AND PREVENTATIVE ACTION (CAPA 0014) HAS BEEN IMPLEMENTED TO INVESTIGATE, CORRECT, AND MONITOR EFFECTIVENESS. AS A RESULT OF THE CAPA INVESTIGATION, IT WAS DECIDED THAT A THICKER 8MM MATERIAL BE USED FOR THE TOP PLATE, REPLACING THE CURRENT 6MM DESIGN. WITH THIS MATERIAL DESIGN CHANGE, IT WAS DETERMINED THE CHANGE TO THE THICKER MATERIAL IMPROVED THE OVERALL STRENGTH OF THE COMPONENT MAKING IT LESS SUSCEPTIBLE TO MATERIAL FATIGUE WHEN EXPOSED TO LOG TERM STRESSES. THE REDESIGNED COMPONENT HAS BEEN INSTALLED ON THE DEVICE, AND THE CHAIR RETURNED TO THE END-USER. THE DHR WAS REVIEWED AND THE DEVICE WAS FOUND TO HAVE MET SPECIFICATION PRIOR TO DISTRIBUTION.

Description of Event or Problem · 1

PERMOBIL AB RECEIVED A REPORT CLAIMING AS THE END-USER WAS IN PROCESS OF PERFORMING A TRANSFER, THE SEATING ALLEGEDLY DETACH AND FELL OFF TO THE SIDE TO THE GROUND. REPORTS INDICATE THE END-USER SUSTAINED MINOR INJURIES THAT DID NOT REQUIRE MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
679256 C500 CORPUS POWERED WHEELCHAIR ITI PERMOBIL AB (PAB) C500 CORPUS N/A

Patients

Seq Age Sex Outcome Treatment
1