FDA Adverse Event Malfunction Summary report: N

IPC HANDPIECE - INDIGO DRILL

MDR report key: 11782859 · Received May 6, 2021

Report

Report Number
1045254-2021-00224
Event Type
Malfunction
Date Received
May 6, 2021
Date of Event
March 30, 2021
Report Date
May 6, 2021
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
UDI-DI
00613994933713
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT ANALYSIS FOUND THAT THE COLLET WAS OVERHEATING. UPON PRE-CHECK, TEMPERATURE AT ONE MINUTE WAS 85.1°F ON MOTOR AND 119.5°F ON COLLET. HAND PIECE WAS TOO DIRTY AND BEARING OF THE COLLET WAS CORRODED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
683149 IPC HANDPIECE - INDIGO DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000 215359015 00613994933713

Patients

Seq Age Sex Outcome Treatment
1