FDA Adverse Event
Malfunction
Summary report: N
IPC HANDPIECE - INDIGO DRILL
MDR report key: 11782859
·
Received May 6, 2021
Report
- Report Number
- 1045254-2021-00224
- Event Type
- Malfunction
- Date Received
- May 6, 2021
- Date of Event
- March 30, 2021
- Report Date
- May 6, 2021
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- HBE
- UDI-DI
- 00613994933713
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
PRODUCT ANALYSIS FOUND THAT THE COLLET WAS OVERHEATING. UPON PRE-CHECK, TEMPERATURE AT ONE MINUTE WAS 85.1°F ON MOTOR AND 119.5°F ON COLLET. HAND PIECE WAS TOO DIRTY AND BEARING OF THE COLLET WAS CORRODED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 683149 | IPC HANDPIECE - INDIGO DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MEDTRONIC XOMED INC. | 1845000 | 215359015 | 00613994933713 |