FDA Adverse Event Malfunction Summary report: N

FAST FX STRUT X-SHORT

MDR report key: 11781896 · Received May 6, 2021

Report

Report Number
1020279-2021-03801
Event Type
Malfunction
Date Received
May 6, 2021
Date of Event
April 12, 2021
Report Date
October 21, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
JDW
UDI-DI
03596010479884
PMA / PMN Number
K994143
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H6: THE ASSOCIATED DEVICE, USED IN TREATMENT, WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THE STATED FAILURE MODE. THE DEVICE IS SEVERELY SCRATCHED UP AND MISSING THE COATING, MAKING IT DIFFICULT TO READ THE RULER. THIS RENDERS THE DEVICE INOPERATIVE. THE DEVICE SHOWS SIGNS OF EXTENSIVE USE. A REVIEW OF COMPLAINT HISTORY DID NOT REVEAL ADDITIONAL COMPLAINTS FOR THE LISTED BATCH. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. WEAR POTENTIAL CAUSES COULD INCLUDE BUT ARE NOT LIMITED TO FRICTION, JOINT TIGHTNESS OR LIFETIME OF DEVICE. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT (3) FAST FX STRUT FOR TSF {} MED ((B)(4)), (6) FAST FX STRUT X-SHORT ((B)(4)), (6) FAST FX STRUT FOR TSF {} SHT ((B)(4)) AND (1) FAST FX STRUT FOR TSF {} LNG ((B)(4)) ARE LOOSING THE BLACK COATING ON THE OUTSIDE MAKING IMPOSSIBLE TO READ THE NUMBERS. SOME ALSO WILL NOT ALLOW THE SHOULDER BOLTS TO THREAD IN. IT IS UNKNOWN WHETHER THE EVENT HAPPENED AND IF THERE WAS A PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682062 FAST FX STRUT X-SHORT PIN, FIXATION, THREADED JDW SMITH & NEPHEW, INC. 71070705 20DSM0178 03596010479884

Patients

Seq Age Sex Outcome Treatment
1