Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: DURING MANUFACTURING OF MATERIAL 221239, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD REVIEW FOR BATCH 1042195 WAS SATISFACTORY AT TIME OF RELEASE PER INTERNAL PROCEDURES. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE BIOBURDEN TESTING. A SAMPLE OF PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. ALL BIOBURDEN TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER BD INTERNAL PROCEDURES. AFFECTED PRODUCT DOES NOT HAVE ANY STERILITY CLAIMS; THE PRODUCT IS TESTED FOR BIOBURDEN PRIOR TO RELEASE TO ENSURE THAT IT CONFORMS TO PRODUCT SPECIFICATIONS. HOWEVER, THIS DOES NOT ENSURE THAT THE END-USER WILL NOT RECEIVE A CONTAMINATED PLATE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 1042195. RETENTION SAMPLES FROM BATCH 1042195 WERE NOT AVAILABLE FOR INSPECTION. NO RETURN SAMPLES OR PHOTOS WERE RECEIVED FOR INVESTIGATION. THIS COMPLAINT CANNOT BE CONFIRMED. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) FUNGAL CONTAMINATION WAS DISCOVERED ON PLATE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTING FUNGAL CONTAMINATION ON 221239 PLATES LOT 1042195.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) FUNGAL CONTAMINATION WAS DISCOVERED ON PLATE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTING FUNGAL CONTAMINATION ON 221239 PLATES LOT 1042195.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
676788 BD BBL TRYPTICASE SOY AGAR WITH 5% SHEEP BLOOD (TSA II) CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON, DICKINSON & CO. (SPARKS) 221239 1042195 10382902212390

Patients

Seq Age Sex Outcome Treatment
1