FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES

MDR report key: 11778710 · Received May 5, 2021

Report

Report Number
1917413-2021-00377
Event Type
Malfunction
Date Received
May 5, 2021
Date of Event
April 12, 2021
Report Date
May 20, 2021
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903678612
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY DRAW TESTING AND NO ISSUES WERE OBSERVED RELATING TO UNDERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED THE BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES EXPERIENCED UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "TUBES DO NOT APPEAR TO BE FILLING COMPLETELY TO THE LINE. SAMPLES ARE GETTING CANCELED FOR THE QUANTITY NOT BEING SUFFICIENT. SEEMS TO BE UNDERFILLING."

Additional Manufacturer Narrative · 1

MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 0281159, MEDICAL DEVICE EXPIRATION DATE: 2020-10-07, DEVICE MANUFACTURE DATE: (B)(6) 2020. MEDICAL DEVICE LOT #: 0100272, MEDICAL DEVICE EXPIRATION DATE: 2022-02-28, DEVICE MANUFACTURE DATE: (B)(6) 2020. MEDICAL DEVICE LOT #: 0258325, MEDICAL DEVICE EXPIRATION DATE: 2022-02-28, DEVICE MANUFACTURE DATE: (B)(6) 2020. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED

Description of Event or Problem · 1

IT WAS REPORTED THE BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES EXPERIENCED UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "TUBES DO NOT APPEAR TO BE FILLING COMPLETELY TO THE LINE. SAMPLES ARE GETTING CANCELED FOR THE QUANTITY NOT BEING SUFFICIENT. SEEMS TO BE UNDERFILLING."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
671864 BD VACUTAINER K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367861 SEE H.10. 50382903678612

Patients

Seq Age Sex Outcome Treatment
1