BD VACUTAINER K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES
Report
- Report Number
- 1917413-2021-00377
- Event Type
- Malfunction
- Date Received
- May 5, 2021
- Date of Event
- April 12, 2021
- Report Date
- May 20, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 50382903678612
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY DRAW TESTING AND NO ISSUES WERE OBSERVED RELATING TO UNDERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.
IT WAS REPORTED THE BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES EXPERIENCED UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "TUBES DO NOT APPEAR TO BE FILLING COMPLETELY TO THE LINE. SAMPLES ARE GETTING CANCELED FOR THE QUANTITY NOT BEING SUFFICIENT. SEEMS TO BE UNDERFILLING."
MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 0281159, MEDICAL DEVICE EXPIRATION DATE: 2020-10-07, DEVICE MANUFACTURE DATE: (B)(6) 2020. MEDICAL DEVICE LOT #: 0100272, MEDICAL DEVICE EXPIRATION DATE: 2022-02-28, DEVICE MANUFACTURE DATE: (B)(6) 2020. MEDICAL DEVICE LOT #: 0258325, MEDICAL DEVICE EXPIRATION DATE: 2022-02-28, DEVICE MANUFACTURE DATE: (B)(6) 2020. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED
IT WAS REPORTED THE BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES EXPERIENCED UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "TUBES DO NOT APPEAR TO BE FILLING COMPLETELY TO THE LINE. SAMPLES ARE GETTING CANCELED FOR THE QUANTITY NOT BEING SUFFICIENT. SEEMS TO BE UNDERFILLING."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 671864 | BD VACUTAINER K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 367861 | SEE H.10. | 50382903678612 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |