FDA Adverse Event Malfunction Summary report: N

ETEST IMIPENEM RELEBACTAM IPR US S30

MDR report key: 11778498 · Received May 5, 2021

Report

Report Number
9615754-2021-00140
Event Type
Malfunction
Date Received
May 5, 2021
Report Date
July 6, 2021
Manufacturer
BIOMERIEUX, SA
Product Code
JWY
UDI-DI
03573026551544
PMA / PMN Number
K191953
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE RESULTS REPORTED LED TO CATEGORY ERROR (RESISTANT RESULTS INSTEAD OF SUSCEPTIBLE) COMPARED TO THE EXPECTED RESULTS MENTIONED IN THE CDC PANEL. THESE ISSUES WERE OBSERVED BY THE CUSTOMER ON THE ETEST® IMIPENEM / RELEBACTAM (IPR) S30, REFERENCE (B)(4), LOT NUMBERS 1007607290 (EXPIRY DATE ON 24 JUN 2021), 1007888870 (EXPIRY DATE ON 09 NOV 2021), 1008390030 (EXPIRY DATE ON 06 AUG 2022) AND 1008598350 (EXPIRY DATE ON 04 DEC 2022). INVESTIGATION: RETAINED SAMPLES FOR THE THREE (3) DIFFERENT LOTS OF ETEST® IPR WERE TESTED DURING THE INVESTIGATION. TESTS WERE PERFORMED USING THE QUALITY CONTROL (QC) STRAINS; PSEUDOMONAS AERUGINOSA ATCC® 27853¿, ESCHERICHIA COLI ATCC® 25922¿, ESCHERICHIA COLI ATCC® 35218¿, KLEBSIELLA PNEUMONIAE ATCC® BAA 1705¿ AND KLEBSIELLA PNEUMONIAE ATCC® 700603¿ ACCORDING TO THE QC PROTOCOL USED FOR THE RELEASE OF EACH LOT. THE RESULTS OBTAINED COMPLIED WITH SPECIFICATIONS FOR THE FIVE (5) STRAINS AND THE THREE (3) ETEST® IPR LOTS TESTED. THE FOURTEEN (14) PSEUDOMONAS AERUGINOSA STRAINS FROM THE CDC PANEL WERE TESTED ON THE THREE (3) LOTS MENTIONED ABOVE, COMPARED TO BMD METHOD. THE MIC OVERESTIMATIONS WERE CONFIRMED. ADDITIONAL TESTS WERE PERFORMED USING 45 ENTEROBACTERIACEAE STRAINS AND 30 PSEUDOMONAS AERUGINOSA STRAINS. THIS STUDY SHOWED NO PERFORMANCE ISSUE WITH ENTEROBACTERIACEAE STRAINS BUT CONFIRMED MIC OVERESTIMATIONS FOR PSEUDOMONAS AERUGINOSA STRAINS. COMPLAINT REVIEW THE INVESTIGATOR REVIEWED THE HISTORICAL COMPLAINTS DATABASE TO FIND ANY SIMILAR COMPLAINTS WHICH WERE PREVIOUSLY REPORTED AGAINST ETEST® IPR INCLUDING REFERENCES (B)(4) (S30 US) AND (B)(4) (S30 WW). APART FROM THE RECORDS RELATED TO THIS ISSUE, NO OTHER COMPLAINTS WERE REGISTERED FOR PRODUCT REFERENCE (B)(4), NOR WERE ANY REGISTERED FOR THE PRODUCT REFERENCE (B)(4). CONCLUSION THE INVESTIGATOR CONFIRMED A PERFORMANCE DEGRADATION OF THE PRODUCT ETEST® IPR LEADING TO AN INCREASE MAJOR ERROR RATE FOR PSEUDOMONAS AERUGINOSA STRAINS. BOTH REFERENCES OF ETEST® IPR, (B)(4) AND (B)(4), ARE IMPACTED BY THIS PERFORMANCE ISSUE. CORRECTIVE AND PREVENTIVE ACTION A FIELD SAFETY CORRECTIVE ACTION (FSCA-5230) HAS BEEN ISSUED 24-JUN-2021, WITH A CUSTOMER LETTER. THIS CUSTOMER LETTER WILL INFORM THE USERS THAT A CONFIRMATION IS NEEDED (USING AN ALTERNATIVE METHOD) ON PSEUDOMONAS AERUGINOSA STRAINS FOR WHICH A RESISTANT OR INTERMEDIATE RESULT HAS BEEN OBTAINED WITH ETEST® IMIPENEM RELEBACTAM ¿ REF. (B)(4) AND (B)(4). FURTHERMORE, A CORRECTIVE ACTION AND PREVENTIVE ACTION PLAN (CAPA) IS IN PROGRESS AT MANUFACTURING SITE LEVEL TO IDENTIFY THE ROOT CAUSE RELATED TO THIS ISSUE AND DETERMINE THE NECESSARY ACTIONS.

Description of Event or Problem · 0

A CUSTOMER IN THE UNITED STATED NOTIFIED BIOMERIEUX OF OBTAINING OVERESTIMATED MIC RESULTS IN ASSOCIATION WITH ETEST® IMIPENEM RELEBACTAM IPR US S30 (REF. 420927, LOT 1008390030) WHILE PERFORMING VALIDATION TESTING WITH SEVEN (7) STRAINS OF PSEUDOMONAS AERUGINOSA INCLUDING PSEUDOMONAS AERUGINOSA ATCC® 27853¿. THE CUSTOMER TESTED THE PSEUDOMONAS AERUGINOSA STRAINS TWICE USING 150MM REMEL PLAIN MUELLER HINTON AGAR PLATES AND 0.45% SALINE SOLUTION WITH A MCFARLAND DENSITY OF BETWEEN 0.5 AND 0.53. OVERESTIMATED IMIPENEM RELEBACTAM MIC RESULTS WERE OBTAINED FOR EACH ISOLATE FOR BOTH INITIAL AND REPEAT TESTING. BIOMERIEUX CUSTOMER SERVICE ADVISED THE CUSTOMER TO USE A 0.85% SALINE SOLUTION AS INSTRUCTED IN THE ETEST® IMIPENEM RELEBACTAM IPR US S30 INSTRUCTIONS FOR USE (IFU). THE CUSTOMER REPEATED TESTING USING A 0.85% SALINE SOLUTION FOLLOWING THE INSTRUCTIONS FOR USE. ALL SEVEN (7) ISOLATES OBTAINED OVERESTIMATED MIC RESULTS WITH FOUR (4) FALSE RESISTANT RESULTS INTERPRETED USING THE CLSI 2021 BREAKPOINTS WITH A SUSCEPTIBLE MIC RESULT =2 AND RESISTANT RESULT >8. THE RESULTS OF THE FALSE RESISTANT RESULTS ARE LISTED BELOW. (B)(6). AS THERE ARE NO PATIENTS ASSOCIATED WITH THESE QUALITY CONTROL STRAINS, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. BIOMERIEUX HAS INITIATED AN INTERNAL INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
674497 ETEST IMIPENEM RELEBACTAM IPR US S30 ETEST® IMIPENEM RELEBACTAM IPR US S30 JWY BIOMERIEUX, SA 1008390030 03573026551544

Patients

Seq Age Sex Outcome Treatment
1