FDA Adverse Event Injury Summary report: N

INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE

MDR report key: 11776167 · Received May 5, 2021

Report

Report Number
1037905-2021-00182
Event Type
Injury
Date Received
May 5, 2021
Date of Event
April 14, 2021
Report Date
June 25, 2021
Manufacturer
COOK ENDOSCOPY
Product Code
PKL
UDI-DI
00827002580107
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION FOLLOW UP REPORT IS BEING SUBMITTED TO CORRECT THE ADVERSE EVENT/PRODUCT PROBLEM TO ADVERSE EVENT AND PRODUCT PROBLEM AND REPORTABLE TYPE TO SERIOUS INJURY. PMA/510(K) # K192697. INVESTIGATION EVALUATION: THE CUSTOMER PROVIDED A PICTURE WITH THE RETURNED DEVICE. THE PICTURE SHOWS THE CLIP OPEN AND STILL ATTACHED TO THE DRIVE WIRE. THE PICTURE CONFIRMS THE CUSTOMER'S COMPLAINT OF PREMATURE DEPLOYMENT WHEN OPENING THE CLIP. USED DEVICE: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED, BECAUSE ALL THE DEVICE COMPONENTS (PARTICULARLY THE CLIP) WERE NOT INCLUDED IN THE RETURN. DURING A FUNCTIONAL TEST, THE HANDLE WAS MANIPULATED, AND THE DRIVE WIRE MOVED FREELY INSIDE THE OUTER SHEATH. THE DEVICE WAS THEN ADVANCED DOWN AN OLYMPUS 2.8 MM CHANNEL ENDOSCOPE IN A SIMULATED UPPER GI POSITION WITH THE TIP IN THE MAXIMUM RETROFLEXED POSITION TO SIMULATE A WORST-CASE POSITION. WITH HANDLE MANIPULATION, THE DRIVE WIRE WAS AGAIN OBSERVED TO MOVE FREELY INSIDE THE OUTER SHEATH. A VISUAL EXAMINATION OF THE DRIVE WIRE, CATHETER ATTACHMENT, AND COIL CATHETER (DISTAL END DEVICE COMPONENTS) SHOWED NO DEFORMITIES OR SIGNS OF DAMAGE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OCCURRENCE WAS NOT OBSERVED. 8 SEALED DEVICES FROM THE SAME LOT: THE SAMPLES WERE FUNCTION TESTED FOR OPEN/CLOSE AND FULL DEPLOYMENT ON SIMULATED TISSUE. THE CLIPS OPENED AND CLOSED, AND DEPLOYED AS INTENDED ON SIMULATED TISSUE. THE COMPLAINT WAS NOT CONFIRMED FOR THE SEALED DEVICES. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: USED DEVICE: A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED BECAUSE THE CONDITION OF THE PRODUCT SAID TO BE INVOLVED PROHIBITED A COMPLETE EVALUATION, THE CLIP HAD BEEN DEPLOYED AND WAS NOT INCLUDED IN THE RETURN. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. SEALED DEVICES: A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED BECAUSE THE RETURNED SEALED DEVICES FUNCTIONED AS INTENDED. A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF DEPLOYMENT DIFFICULTY FOR INSTINCT PLUS CLIPS. THE SYSTEM PREPARATION SECTION OF THE INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP. FAILURE TO KEEP ENDOSCOPE AS STRAIGHT AS POSSIBLE WHEN INSERTING DEVICE CAN RESULT IN DIFFICULT PASSAGE.¿ FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

PMA/510(K) # K192697. INVESTIGATION EVALUATION: THE CUSTOMER PROVIDED A PICTURE WITH THE RETURNED DEVICE. THE PICTURE SHOWS THE CLIP OPEN AND STILL ATTACHED TO THE DRIVE WIRE. THE PICTURE CONFIRMS THE CUSTOMER'S COMPLAINT OF PREMATURE DEPLOYMENT WHEN OPENING THE CLIP. USED DEVICE: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED, BECAUSE ALL THE DEVICE COMPONENTS (PARTICULARLY THE CLIP) WERE NOT INCLUDED IN THE RETURN. DURING A FUNCTIONAL TEST, THE HANDLE WAS MANIPULATED, AND THE DRIVE WIRE MOVED FREELY INSIDE THE OUTER SHEATH. THE DEVICE WAS THEN ADVANCED DOWN AN OLYMPUS 2.8 MM CHANNEL ENDOSCOPE IN A SIMULATED UPPER GI POSITION WITH THE TIP IN THE MAXIMUM RETROFLEXED POSITION TO SIMULATE A WORST-CASE POSITION. WITH HANDLE MANIPULATION, THE DRIVE WIRE WAS AGAIN OBSERVED TO MOVE FREELY INSIDE THE OUTER SHEATH. A VISUAL EXAMINATION OF THE DRIVE WIRE, CATHETER ATTACHMENT, AND COIL CATHETER (DISTAL END DEVICE COMPONENTS) SHOWED NO DEFORMITIES OR SIGNS OF DAMAGE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OCCURRENCE WAS NOT OBSERVED. 8 SEALED DEVICES FROM THE SAME LOT: THE SAMPLES WERE FUNCTION TESTED FOR OPEN/CLOSE AND FULL DEPLOYMENT ON SIMULATED TISSUE. THE CLIPS OPENED AND CLOSED, AND DEPLOYED AS INTENDED ON SIMULATED TISSUE. THE COMPLAINT WAS NOT CONFIRMED FOR THE SEALED DEVICES. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: USED DEVICE: A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED BECAUSE THE CONDITION OF THE PRODUCT SAID TO BE INVOLVED PROHIBITED A COMPLETE EVALUATION, THE CLIP HAD BEEN DEPLOYED AND WAS NOT INCLUDED IN THE RETURN. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. SEALED DEVICES: A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED BECAUSE THE RETURNED SEALED DEVICES FUNCTIONED AS INTENDED. A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF DEPLOYMENT DIFFICULTY FOR INSTINCT PLUS CLIPS. THE SYSTEM PREPARATION SECTION OF THE INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP. FAILURE TO KEEP ENDOSCOPE AS STRAIGHT AS POSSIBLE WHEN INSERTING DEVICE CAN RESULT IN DIFFICULT PASSAGE.¿ FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

PMA/510(K) # K192697 INVESTIGATION EVALUATION: THE CUSTOMER PROVIDED A PICTURE WITH THE RETURNED DEVICE. THE PICTURE SHOWS THE CLIP OPEN AND STILL ATTACHED TO THE DRIVE WIRE. THE PICTURE CONFIRMS THE CUSTOMER'S COMPLAINT OF PREMATURE DEPLOYMENT WHEN OPENING THE CLIP. OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED COULD NOT CONFIRM THE REPORT AS IT WAS DESCRIBED, BECAUSE ALL THE DEVICE COMPONENTS (PARTICULARLY THE CLIP) WERE NOT INCLUDED IN THE RETURN. DURING A FUNCTIONAL TEST, THE HANDLE WAS MANIPULATED, AND THE DRIVE WIRE MOVED FREELY INSIDE THE OUTER SHEATH. THE DEVICE WAS THEN ADVANCED DOWN AN OLYMPUS 2.8 MM CHANNEL ENDOSCOPE IN A SIMULATED UPPER GI POSITION WITH THE TIP IN THE MAXIMUM RETROFLEXED POSITION TO SIMULATE A WORST-CASE POSITION. WITH HANDLE MANIPULATION, THE DRIVE WIRE WAS AGAIN OBSERVED TO MOVE FREELY INSIDE THE OUTER SHEATH. A VISUAL EXAMINATION OF THE DRIVE WIRE, CATHETER ATTACHMENT, AND COIL CATHETER (DISTAL END DEVICE COMPONENTS) SHOWED NO DEFORMITIES OR SIGNS OF DAMAGE. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED BECAUSE THE CONDITION OF THE PRODUCT SAID TO BE INVOLVED PROHIBITED A COMPLETE EVALUATION, THE CLIP HAD BEEN DEPLOYED AND WAS NOT INCLUDED IN THE RETURN. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. THE SYSTEM PREPARATION SECTION OF THE INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP. FAILURE TO KEEP ENDOSCOPE AS STRAIGHT AS POSSIBLE WHEN INSERTING DEVICE CAN RESULT IN DIFFICULT PASSAGE.¿ FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE, THE PHYSICIAN USED TWO (2) COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. WHEN THE CLIP WAS OPENED IT ACCIDENTLY DEPLOYED OFF THE SHEATH. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A HEMOSPRAY DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
672262 INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE PKL, LIGATOR, HEMORRHOIDAL PKL COOK ENDOSCOPY G58010 W4416641 00827002580107

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention OLYMPUS ENDOSCOPE, GIF H 190