NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2021-00530
- Event Type
- Malfunction
- Date Received
- May 4, 2021
- Date of Event
- April 6, 2021
- Report Date
- April 6, 2021
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
PS366896 METHOD: THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTRE IN CALIFORNIA WHERE IT WAS TESTED BY A TRAINED F&P SERVICE TECHNICIAN. THE UNIT WAS SERVICED AND PERFORMANCE TESTED. RESULTS: THE PERFORMANCE TEST REVEALED THAT THE MANOMETER NEEDLE WAS STUCK. CONCLUSION: THE LIKELY CAUSE OF THE REPORTED EVENT IS PHYSICAL DAMAGE DUE TO IMPACT. THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT. IN ADDITION THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF". EACH NEOPUFF UNIT IS ASSEMBLED AND 100% TESTED IN THE PRODUCTION LINE TO VERIFY THAT EACH UNIT CONFORMS TO CRITICAL PRODUCT SPECIFICATIONS. ANY UNIT THAT FAILS IS REJECTED. THE SUBJECT NEOPUFF WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT TIME OF PRODUCTION.
A HEALTHCARE FACILITY IN NEW YORK REPORTED THAT A RD900 NEOPUFF INFANT RESUSCITATOR MANOMETER WAS FAULTY. THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE(F&P) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT A RD900 NEOPUFF INFANT RESUSCITATOR MANOMETER WAS FAULTY. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670798 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | RD900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |