FDA Adverse Event Injury Summary report: N

ROHO AGILITY CARBON MID CONTOUR BACK SUPPORT

MDR report key: 11772723 · Received May 4, 2021

Report

Report Number
3008630266-2021-00008
Event Type
Injury
Date Received
May 4, 2021
Date of Event
February 5, 2021
Report Date
April 5, 2021
Manufacturer
ROHO, INC
Product Code
KNO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

ROHO, INC, RECEIVED NOTIFICATION OF THE ADVERSE EVENT THROUGH A LETTER FROM A LAW FIRM. THE FULL EXTENT OF THE INJURIES ARE UNKNOWN. PICTURES OF THE BACK SUPPORT WERE RECEIVED SHOWING THAT A BOLT BROKE, BUT THE BACK SUPPORT HAS NOT BEEN RETURNED TO ROHO, INC. FOR AN EVALUATION. NO DETERMINATION OF THE ALLEGED FAILURE CAN BE MADE. THE FINAL ACCEPTANCE RECORDS FOR THE CUSHION WERE REVIEWED AND IT WAS CONFIRMED THAT THE BACK SUPPORT WAS BUILT ACCORDING TO THE DEVICE MASTER RECORD. FURTHER REQUESTS FOR INFORMATION HAS BEEN SUBMITTED TO THE LAW FIRM AND IF ADDITIONAL INFORMATION IS PROVIDED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

USER ALLEGES THAT DURING NORMAL USE THE BACK REST SNAPPED BACKWARDS CAUSING THE USER TO TIP OVER BACKWARDS AND KNEE HIMSELF IN THE FACE, BREAKING HIS NOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666124 ROHO AGILITY CARBON MID CONTOUR BACK SUPPORT WHEELCHAIR BACK SUPPORT KNO ROHO, INC AG3-1710A-2TCB

Patients

Seq Age Sex Outcome Treatment
1 Other