FDA Adverse Event Malfunction Summary report: N

CHLORAPREP ONE STEP HI LITE ORANGE

MDR report key: 11771868 · Received May 4, 2021

Report

Report Number
3004932373-2021-00230
Event Type
Malfunction
Date Received
May 4, 2021
Date of Event
April 12, 2021
Report Date
July 27, 2021
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
KXG
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

FACILITY DID NOT PROVIDE PHOTOS/SAMPLES TO AID IN OUR QUALITY ENGINEER¿S INVESTIGATION. A DEVICE HISTORY RECORD WAS REVIEWED, AND NO NON-CONFORMANCES WAS NOTED DURING THE MANUFACTURING OF THIS LOT. THE ROOT CAUSE IS ATTRIBUTED TO THE EQUIPMENT STATION FOR THE END CAP PLACEMENT UNTO THE APPLICATOR BODY. CORRECTIVE ACTIONS WERE INITIATED WHICH LED TO BD CONDUCTING A VOLUNTARY RECALL ON CERTAIN LOTS OF THE CHLORAPREP HI- LITE ORANGE 26 ML APPLICATOR. BD HAS CONFIRMED THAT SOME OF THE PRODUCT, WHICH INCLUDED THIS LOT, HAD AN APPLICATOR END CAP THAT WAS IMPROPERLY SECURED DURING THE MANUFACTURING PROCESS WHICH RESULTED IN BROKEN GLASS DROPPING OUT OF THE APPLICATOR. H3 OTHER TEXT : SEE NARRATIVE BELOW.

Description of Event or Problem · 0

IT WAS REPORTED THE APPLICATORS CRACKED AND GLASS WENT ALL ON THE FLOOR AND PATIENT. PER EMAIL:THE FOLLOWING COMPLAINT WAS RECEIVED TODAY OVER THE PHONE: - - MATERIAL INFO: - CATALOG: 930815. - LOT: 0323213. - SAMPLES UNAVAILABLE. - DESCRIPTION OF EVENT: 3 TOTAL CHLORAPREPS CRACKED AND GLASS WENT ALL OVER PATIENT/FLOOR. 2 EVENTS ON 12TH (PATIENT/FLOOR), AND 1 EVENT YESTERDAY (B)(6) 2021 GLASS WENT ON FLOOR. - KNOWN DATES OF EVENTS: 2 ON (B)(6) 2021; 1 ON (B)(6) 2021. - ADVERSE EVENT: NONE; NO PATIENT HARM- ABLE TO SAFELY REMOVE AND CLEAN OFF PATIENT.

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

MATERIAL NO.: 930815, BATCH NO.: 0323213. IT WAS REPORTED THE APPLICATORS CRACKED AND GLASS WENT ALL ON THE FLOOR AND PATIENT. PER EMAIL:THE FOLLOWING COMPLAINT WAS RECEIVED TODAY OVER THE PHONE: - MATERIAL INFO: CATALOG: 930815 . LOT: 0323213. SAMPLES UNAVAILABLE. DESCRIPTION OF EVENT: 3 TOTAL CHLORAPREPS CRACKED AND GLASS WENT ALL OVER PATIENT/FLOOR. 2 EVENTS ON 12TH (PATIENT/FLOOR), AND 1 EVENT YESTERDAY (B)(6) 2021 GLASS WENT ON FLOOR. KNOWN DATES OF EVENTS: 2 ON (B)(6) 2021. ADVERSE EVENT: NONE; NO PATIENT HARM- ABLE TO SAFELY REMOVE AND CLEAN OFF PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
667420 CHLORAPREP ONE STEP HI LITE ORANGE 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL KXG CAREFUSION 213, LLC 0113 0323213

Patients

Seq Age Sex Outcome Treatment
1 Other