FDA Adverse Event Malfunction Summary report: N

FS DISPOSABLE INTERFACE

MDR report key: 11771114 · Received May 4, 2021

Report

Report Number
3006695864-2021-07596
Event Type
Malfunction
Date Received
May 4, 2021
Date of Event
April 9, 2021
Report Date
July 14, 2021
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
GEX
UDI-DI
15050474534688
PMA / PMN Number
K060372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

H3 DEVICE EVALUATED BY MANUFACTURER; YES. DEVICE EVALUATION: AT THE TIME OF THE INVESTIGATION, PRODUCT WAS NOT AVAILABLE FOR EVALUATION. THEREFORE, NO TESTING COULD BE PERFORMED. THE REPORTED EVENT CANNOT BE CONFIRMED. MANUFACTURING RECORD EVALUATION: THE MANUFACTURING RECORDS FOR THE PATIENT INTERFACE WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. A SEARCH REVEALED THAT 2 ADDITIONAL COMPLAINTS FOR THIS LOT NUMBER HAVE BEEN RECEIVED. CONCLUSION: BASED ON THE INFORMATION OBTAINED, PRODUCT MALFUNCTION AND PRODUCT DEFICIENCY CANNOT BE CONFIRMED. NO FURTHER INVESTIGATION IS REQUIRED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

THE PATIENT INTERFACE (PI) SUCTION RING MAY LOSE SUCTION DURING A PROCEDURE. LABEL COPY STATES CORNEAL FIXATION VACUUM LOSS CAN OCCUR. THERE ARE SEVERAL FACTORS THAT MAY CONTRIBUTE TO SUCTION ISSUES SUCH AS DOCTOR¿S TECHNIQUE IN APPLYING THE SUCTION RING TO THE CORNEA, DOCTOR¿S TECHNIQUE IN SQUEEZING THE PI CLIP TO SECURE THE SUCTION RING TO THE PI CONE AND PATIENT ANATOMY AFFECTING THE INTERFACE BETWEEN THE PATIENT¿S CORNEA AND THE SUCTION RING. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT PATIENT INTERFACE HAD SUCTION LOSS DURING LASER FIRING. THERE WAS NO PATIENT INJURY OR SURGICAL INTERVENTION REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
667741 FS DISPOSABLE INTERFACE POWERED LASER SURGICAL INSTRUMENT GEX AMO MANUFACTURING USA, LLC 590106AN 60280847 15050474534688

Patients

Seq Age Sex Outcome Treatment
1 32 YR LASER SERIAL NUMBER: (B)(4)