FS DISPOSABLE INTERFACE
Report
- Report Number
- 3006695864-2021-07596
- Event Type
- Malfunction
- Date Received
- May 4, 2021
- Date of Event
- April 9, 2021
- Report Date
- July 14, 2021
- Manufacturer
- AMO MANUFACTURING USA, LLC
- Product Code
- GEX
- UDI-DI
- 15050474534688
- PMA / PMN Number
- K060372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- PHYSICIAN
Narratives
H3 DEVICE EVALUATED BY MANUFACTURER; YES. DEVICE EVALUATION: AT THE TIME OF THE INVESTIGATION, PRODUCT WAS NOT AVAILABLE FOR EVALUATION. THEREFORE, NO TESTING COULD BE PERFORMED. THE REPORTED EVENT CANNOT BE CONFIRMED. MANUFACTURING RECORD EVALUATION: THE MANUFACTURING RECORDS FOR THE PATIENT INTERFACE WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. A SEARCH REVEALED THAT 2 ADDITIONAL COMPLAINTS FOR THIS LOT NUMBER HAVE BEEN RECEIVED. CONCLUSION: BASED ON THE INFORMATION OBTAINED, PRODUCT MALFUNCTION AND PRODUCT DEFICIENCY CANNOT BE CONFIRMED. NO FURTHER INVESTIGATION IS REQUIRED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
THE PATIENT INTERFACE (PI) SUCTION RING MAY LOSE SUCTION DURING A PROCEDURE. LABEL COPY STATES CORNEAL FIXATION VACUUM LOSS CAN OCCUR. THERE ARE SEVERAL FACTORS THAT MAY CONTRIBUTE TO SUCTION ISSUES SUCH AS DOCTOR¿S TECHNIQUE IN APPLYING THE SUCTION RING TO THE CORNEA, DOCTOR¿S TECHNIQUE IN SQUEEZING THE PI CLIP TO SECURE THE SUCTION RING TO THE PI CONE AND PATIENT ANATOMY AFFECTING THE INTERFACE BETWEEN THE PATIENT¿S CORNEA AND THE SUCTION RING. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED BY THE CUSTOMER THAT PATIENT INTERFACE HAD SUCTION LOSS DURING LASER FIRING. THERE WAS NO PATIENT INJURY OR SURGICAL INTERVENTION REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 667741 | FS DISPOSABLE INTERFACE | POWERED LASER SURGICAL INSTRUMENT | GEX | AMO MANUFACTURING USA, LLC | 590106AN | 60280847 | 15050474534688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | LASER SERIAL NUMBER: (B)(4) |