FDA Adverse Event Malfunction Summary report: N

INOMAX DSIR PLUS (DELIVERY SYSTEM)

MDR report key: 11769591 · Received May 4, 2021

Report

Report Number
3004531588-2021-00111
Event Type
Malfunction
Date Received
May 4, 2021
Date of Event
March 26, 2021
Report Date
May 4, 2021
Manufacturer
MALLINCKRODT MANUFACTURING LLC.
Product Code
MRN
UDI-DI
00867538000106
PMA / PMN Number
K061901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

REPORTABLE MALFUNCTION WITH INOMAX DSIR PLUS (B)(4) WAS DOCUMENTED AND INVESTIGATED UNDER (B)(4). THE REGIONAL SERVICE CENTER (RSC) REPRODUCED THE REPORTED FAILED NO SENSOR ALARM UPON BOOT-UP OF THE RETURNED DEVICE. A LOW AND HIGH NO SENSOR CALIBRATION WAS SUCCESSFULLY PERFORMED AND THE FAILED NO SENSOR ALARM WAS CLEARED. A REVIEW OF THE DEVICE'S SERVICE LOG REVEALED THAT A DELIVERY STOPPED ALARM OCCURRED AS THE MONITORED NITRIC OXIDE (NO) VALUE (I.E., ACTUAL VALUE: 101 PPM) WAS GREATER THAN THE ABSOLUTE MAX OF 100 PPM FOR 12 CONSECUTIVE SECONDS. AFTERWARDS, A SERVICE LOG ENTRY FOR A FAILED LOW NO CELL CALIBRATION DUE TO THE LOW POINTS BEING GREATER THAN THE MAXIMUM VALUE (MAX: 655 COUNTS; ACTUAL VALUE: 1023 COUNTS) WAS OBSERVED. ALTHOUGH IDENTIFIED IN THE SERVICE LOG, THE RSC DID NOT EXPERIENCE A DELIVERY STOPPED ALARM DURING TESTING. THE ROOT CAUSE FOR THIS OCCURRENCE WAS LIKELY DUE TO A MALFUNCTIONING NO SENSOR. AS A RESULT, THE DEVICE'S NO SENSOR WAS REPLACED. TRENDS WERE REVIEWED FOR THIS REPORTABLE CONDITION AND DETERMINED TO BE WITHIN THE ESTABLISHED CONTROL LIMITS. NO FURTHER ACTION IS REQUIRED AT THIS TIME. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS. AS A RESULT, THE DEVICE WAS PLACED BACK INTO CIRCULATION FOR CUSTOMER USE.

Description of Event or Problem · 1

A CUSTOMER LOCATED IN THE UNITED STATES CONTACTED MALLINCKRODT ON 28-MAR-2021 TO REPORT SECOND-HAND THAT A FAILED NITRIC OXIDE (NO) SENSOR WAS EXPERIENCED WITH THEIR INOMAX DSIR PLUS DEVICE. AS A RESULT, THE DEVICE WAS RETURNED TO MALLINCKRODT FOR INVESTIGATION. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING A ROUTINE SERVICE LOG REVIEW FOR A DEVICE LOCATED IN THE UNITED STATES, A MALLINCKRODT EMPLOYEE DISCOVERED THAT A DELIVERY STOPPED ALARM HAD OCCURRED DUE TO A NO CONCENTRATION GREATER THAN 100 PPM FOLLOWED BY A FAILED LOW NO SENSOR CALIBRATION. THIS SERVICE LOG FINDING MEETS THE CRITERIA OF A REPORTABLE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
665441 INOMAX DSIR PLUS (DELIVERY SYSTEM) APPARATUS MRN MALLINCKRODT MANUFACTURING LLC. 10007 00867538000106

Patients

Seq Age Sex Outcome Treatment
1