NITROSPRAY PLUS
Report
- Report Number
- 2511556-2020-00002
- Event Type
- Injury
- Date Received
- May 3, 2021
- Date of Event
- December 1, 2020
- Report Date
- December 2, 2020
- Manufacturer
- PREMIER DENTAL PRODUCTS CO.
- Product Code
- GEH
- UDI-DI
- D6911006060
- PMA / PMN Number
- K970992
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 501
Narratives
DURING THE INTERVIEW WITH THE OFFICE, IT WAS DISCLOSED THAT A SPRAY TIP WAS NOT APPLIED TO THE NITROSPRAY PLUS NOZZLE. THE INSTRUCTIONS FOR USE WERE NOT FOLLOWED. SUBMITTING REPORT AGAIN TO CORRECT THE ERROR OF SUMMARY REPORT BOX BEING CHECKED.
DUE TO USER ERROR, TWO PATIENTS HAD BLISTERING AFTER TREATMENT. PATIENTS WERE MONITORED FOR HEALING, BOTH BURNS WERE NOT SERIOUS AND DID NOT REQUIRE MEDICAL TREATMENT.
DURING THE INTERVIEW WITH THE OFFICE, IT WAS DISCLOSED THAT A SPRAY TIP WAS NOT APPLIED TO THE NITROSPRAY PLUS NOZZLE. THE INSTRUCTIONS FOR USE WERE NOT FOLLOWED. ORIGINAL MFR REPORT NUMBER 2511556-2020-00001 WAS FILED.
DUE TO USER ERROR, TWO PATIENTS HAD BLISTERING AFTER TREATMENT. PATIENTS WERE MONITORED FOR HEALING, BOTH BURNS WERE NOT SERIOUS AND DID NOT REQUIRE MEDICAL TREATMENT. THIS IS THE MDR REPORT FOR THE 2ND PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 657319 | NITROSPRAY PLUS | CRYOSURGICAL UNIT | GEH | PREMIER DENTAL PRODUCTS CO. | 1006060 | 14623 | D6911006060 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |