FDA Adverse Event Injury Summary report: N

ACL TOP 750 CTS

MDR report key: 11759548 · Received May 3, 2021

Report

Report Number
1217183-2021-00001
Event Type
Injury
Date Received
May 3, 2021
Date of Event
April 19, 2021
Report Date
May 3, 2021
Manufacturer
INSTRUMENTATION LABORATORY CO.
Product Code
GKP
UDI-DI
08426950784067
PMA / PMN Number
K150877
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AS PART OF THE INVESTIGATION, SERVICE WAS DISPATCHED FOR ACL TOP 750 CTS ANALYZER AND INSTRUMENTATION LABORATORY (IL) REVIEWED THE ONLINE HELP DOCUMENTATION (OPERATOR'S MANUAL) AND CONFIRMED THAT THERE ARE APPROPRIATE INSTRUCTIONS AND WARNINGS REGARDING THE CLEANING OF THE PIERCER PROBE ASSEMBLY. THE INSTRUCTIONS INCLUDE A WARNING NOT TO PLACE A FINGER UNDER THE ACL TOP 750 CTS SPECIMEN PIERCER SINCE IT IS VERY SHARP. THE ACL TOP 750 CTS PERFORMED AS INTENDED WITH NO MALFUNCTION AND ITS LABELING PROVIDES APPROPRIATE WARNINGS TO THE USER REGARDING THE HAZARD OF THE SAMPLE PIERCER. THEREFORE, NO REMEDIAL ACTION IS INDICATED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT A CUSTOMER WAS PERFORMING THE MANUAL NEEDLE CLEANING PROCEDURE OF THE PIERCER PROBE ASSEMBLY ON THE ACL TOP 750 CTS ANALYZER AND PUNCTURED HIS FINGER. FOLLOWING HOSPITAL PROTOCOL, TREATMENT WAS PROVIDED IN THE EMERGENCY ROOM OF THE HOSPITAL. MEDICAMENT AND BANDAGE WERE PROVIDED TO THE INJURED FINGER. THERE WAS NO KNOWN ADVERSE EFFECT FROM THE PUNCTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
656209 ACL TOP 750 CTS COAGULATION ANALYZER GKP INSTRUMENTATION LABORATORY CO. 2800-55 08426950784067

Patients

Seq Age Sex Outcome Treatment
1 Other