FDA Adverse Event Malfunction Summary report: Y

MIDWEST TRADITION HANDPIECE

MDR report key: 11758366 · Received April 30, 2021

Report

Report Number
9614977-2021-00012
Event Type
Malfunction
Date Received
April 30, 2021
Report Date
June 9, 2021
Manufacturer
SIRONA DENTAL SYSTEMS GMBH
Product Code
EFB
PMA / PMN Number
K131319
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE INFORMATION IN FIELD D3 WAS INADVERTENTLY MISSED DURING THE INITIAL SUBMISSION. THIS SUBMISSION IS TO CORRECT THE REPORT TO INCLUDE THIS ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 1

ADDITIONAL SERIAL NUMBERS INCLUDED IN THIS REPORT: (B)(4). 5 OF 13 DEVICES WERE RETURNED FOR EVALUATION. WE ARE AWAITING THE RETURN OF 8 DEVICES. EVALUATION OF 4 DEVICES FOUND NORMAL CHUCK WEAR AND THE TURBINES NEEDED TO BE REPLACED. THE DEVICES WERE REPAIRED AND RETURNED TO THE CUSTOMERS. EVALUATION OF 1 DEVICE FOUND CHUCK WEAR AND LACK OF MAINTENANCE. THE DEVICE WAS CLEANED OF DEBRIS AND THE TURBINE WAS REPLACED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

THIS REPORT SUMMARIZES 13 MALFUNCTION EVENTS WHERE A MIDWEST TRADITION HANDPIECE WOULD NOT HOLD BURS. NO INJURY RESULTED IN ANY OF THE EVENTS.