FDA Adverse Event
Malfunction
Summary report: Y
MIDWEST TRADITION HANDPIECE
MDR report key: 11758356
·
Received April 30, 2021
Report
- Report Number
- 9614977-2021-00007
- Event Type
- Malfunction
- Date Received
- April 30, 2021
- Report Date
- June 9, 2021
- Manufacturer
- SIRONA DENTAL SYSTEMS GMBH
- Product Code
- EFB
- PMA / PMN Number
- K963050
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THE INFORMATION IN FIELD D3 WAS INADVERTENTLY MISSED DURING THE INITIAL SUBMISSION. THIS SUBMISSION IS TO CORRECT THE REPORT TO INCLUDE THIS ADDITIONAL INFORMATION.
Additional Manufacturer Narrative · 1
1 OF 1 DEVICE WAS RETURNED FOR EVALUATION. EVALUATION OF 1 DEVICE FOUND NORMAL CHUCK WEAR AND THE TURBINE NEEDED TO BE REPLACED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT WHERE A MIDWEST TRADITION HANDPIECE WOULD NOT HOLD BURS. NO INJURY RESULTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 649716 | MIDWEST TRADITION HANDPIECE | HANDPIECE, AIR-POWERED, DENTAL | EFB | SIRONA DENTAL SYSTEMS GMBH | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |