FDA Adverse Event Malfunction Summary report: Y

DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D100

MDR report key: 11758080 · Received April 30, 2021

Report

Report Number
3001556265-2021-00004
Event Type
Malfunction
Date Received
April 30, 2021
Date of Event
April 8, 2021
Report Date
April 30, 2021
Manufacturer
AGFA N.V.
Product Code
IZL
PMA / PMN Number
K103597
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT. THE CUSTOMER REPORTED AN EVENT OF UNINTENDED MOVMENT OF A DX-D 100 SYSTEM. THE CUSTOMER DESCRIBED THE SYSTEM WAS DRIVING ERRATICALLY. THE RESPONDING SERVICE DEALER ENGINEER STATED THE CUSTOMER REPORTED THE DX-D 100 SYSTEM WOULD SPEED UP AND THEN TURN SHARPLY BEFORE STOPPING. THE DX-D 100 CONTINUED WHEN THE DEAD MAN SWITCH WAS NOT ENGAGED. THERE WERE NO REPORTED ERROR CODES OR BEEPS FROM THE SYSTEM. THE DEALER SERVICE ENGINEER CHECKED AND ADJUSTED THE DRIVE GAUGES AS PER THE SERVICE MANUAL. THEN ENGINEER ALSO REPLACED THE GAUGES, DEADMAN SWITCHES, AND DMC BOARD. AFTER A SYSTEM CHECK THE ISSUE REOCCURED. THE DEALER SERVICE ENGINEER THEN REPLACED THE MOTORS AND ENCODER CABLES. THE SYSTEM WAS TESTED AND IS PERFORMING AS INTENDED. THERE HAVE BEEN NO REPORTS OF ADDITIONAL EVENTS. THERE WAS NO REPORTED HARM TO PATIENTS OR USERS AND THERE ARE NO ADDITIONAL ACTIONS FOR THESE EVENTS.