FDA Adverse Event Malfunction Summary report: N

API STAPH 25STRIPS+25MEDIA

MDR report key: 11755693 · Received April 30, 2021

Report

Report Number
9615754-2021-00137
Event Type
Malfunction
Date Received
April 30, 2021
Report Date
June 14, 2021
Manufacturer
BIOMÉRIEUX, SA
Product Code
JWX
UDI-DI
03573026048259
PMA / PMN Number
C1 EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION CONFIRMED THAT THERE IS AN ERROR IN THE INSTRUCTIONS FOR USE (IFU) VERSION N. INVESTIGATOR CONFIRMED THAT THE EXPECTED ADH RESULT FOR THE STRAIN STAPHYLOCOCCUS CAPITIS ATCC® 35661¿ IS POSITIVE. THE ROOT CAUSE OF THE ISSUE IS A HUMAN TRANSCRIPTION ERROR DURING THE UPDATE OF VERSION M TO N OF THE IFU. THIS ERROR IS A DOCUMENTATION ERROR, AND NO DIRECT IMPACT ON THE PRODUCT PERFORMANCE. IN THE FRAMEWORK OF THE INVESTIGATION, THE VERSION N OF THE IFU WAS ENTIRELY REVIEWED, AND NO OTHER MISTAKE WAS DETECTED. AS CORRECTIVE ACTION, THE IFU WAS CORRECTED, AND THE FIRST BATCH BEING MANUFACTURED WITH THE IFU VERSION O IS BATCH 1008701410. A COMMUNICATION WAS SENT TO ALL CUSTOMERS USING API STAPH REFERENCE 20500 TO INFORM THEM ABOUT THE CORRECTION AND UPDATE OF THE IFU TO VERSION O.

Description of Event or Problem · 0

A CUSTOMER FROM (B)(6) REPORTED HE OBSERVED A DISCREPANCY BETWEEN THE INSTRUCTIONS FOR USE (IFU) VERSION M AND N FOR THE ADH TEST OF QUALITY CONTROL STRAIN S.CAPITIS FOR THE PRODUCT API® STAPH 25STRIPS+25MEDIA (REF. 20500). WHEN PERFORMING QUALITY ASSURANCE TESTING, THE CUSTOMER GOT A POSITIVE RESULT FOR ADH TEST USING STAPHYLOCOCCUS CAPITIS ATCC® 35661¿, AND THEIR RESULT FOR ADH TEST DID NOT CONFORM WITH THE PROFILE INCLUDED IN THE PACKAGE INSERT 07468 VERSION N. THE VERSION ¿07468 M - EN - 2017/03¿ HAS THE EXPECTED ADH TEST RESULT IN THE PROFILE FOR STAPHYLOCOCCUS CAPITIS LISTED AS POSITIVE WHILE THE NEWEST VERSION ¿07468 - N - 2019/05 ¿ EN¿ HAS THE EXPECTED ADH TEST RESULT IN THE PROFILE FOR STAPHYLOCOCCUS CAPITIS LISTED AS NEGATIVE. THE ADH TEST RESULT IN THE PROFILE FOR STAPHYLOCOCCUS CAPITIS WAS MODIFIED BY MISTAKE IN THE VERSION N OF THE IFU. THE CUSTOMER REPORTED THEY PERFORM QUALITY CONTROL WITH S.CAPITIS ATCC® 35661¿ ON EVERY LOT RECEIVED AND ALWAYS OBSERVED A POSITIVE RESULT FOR THE ADH TEST WITH S.CAPITIS STRAIN. THE CUSTOMER ALSO TESTS STAPHYLOCOCCUS XYLOSUS ATCC® 700404¿ AND STAPHYLOCOCCUS LENTUS ATCC® 700403¿ WHEN RECEIVING A NEW LOT. THE INFORMATION FOR THESE STRAINS WERE NOT CHANGED IN THE IFU. THERE IS NO INDICATION FROM THE CUSTOMER THAT THIS DOCUMENTATION ERROR LED TO A DELAY OF REPORTING RESULTS. A BIOMÉRIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
654406 API STAPH 25STRIPS+25MEDIA API® STAPH 25STRIPS+25MEDIA JWX BIOMÉRIEUX, SA 20500 03573026048259

Patients

Seq Age Sex Outcome Treatment
1