FDA Adverse Event Malfunction Summary report: N

COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER

MDR report key: 11755620 · Received April 30, 2021

Report

Report Number
1820334-2021-01264
Event Type
Malfunction
Date Received
April 30, 2021
Date of Event
March 31, 2021
Report Date
July 23, 2021
Manufacturer
COOK INC
Product Code
LKF
UDI-DI
00827002178700
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNCHANGED, OR UNAVAILABLE. EVENT DESCRIPTION: AS REPORTED, A COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER RUPTURED IN THE PATIENT WHEN INFLATED WITH 1.3ML SALINE. THE DEVICE WAS NOT TESTED PRIOR TO USE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS A RESULT OF THIS OCCURRENCE. INVESTIGATION ¿ EVALUATION A VISUAL INSPECTION OF THE RETURNED DEVICE WAS CONDUCTED. A DOCUMENT BASED INVESTIGATION WAS ALSO PERFORMED INCLUDING A REVIEW OF COMPLAINT HISTORY, DEVICE HISTORY RECORD, THE INSTRUCTIONS FOR USE, AND QUALITY CONTROL DATA. ONE COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER WAS RETURNED FOR INVESTIGATION. INSPECTION OF THE RETURNED DEVICE CONFIRMED THAT THE BALLOON WAS RUPTURED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO NON-CONFORMANCES RELATED TO THE REPORTED FAILURE MODE. A REVIEW OF COMPLAINT HISTORY RECORDS SHOWS NO OTHER COMPLAINTS ASSOCIATED WITH THE COMPLAINT DEVICE LOT. BECAUSE THERE WERE NO RELATED NON-CONFORMANCES, ADEQUATE INSPECTION ACTIVITIES HAVE BEEN ESTABLISHED, THERE IS OBJECTIVE EVIDENCE THAT THE DHR WAS FULLY EXECUTED, AND NO OTHER LOT RELATED COMPLAINTS THAT HAVE BEEN RECEIVED FROM THE FIELD, IT WAS CONCLUDED THAT THERE IS NO EVIDENCE THAT NONCONFORMING PRODUCT EXISTS IN HOUSE OR IN THE FIELD. A REVIEW OF RELEVANT MANUFACTURING DOCUMENTS WAS CONDUCTED. IT WAS CONCLUDED THAT THE DEVICE ASPECT IN QUESTION WAS FUNCTIONALLY INSPECTED BY QUALITY CONTROL AND NO NOTABLE GAPS IN PRODUCTION OR PROCESSING CONTROLS WERE NOTED. THERE IS NO INDICATION THAT A DESIGN OR PROCESS RELATED FAILURE MODE CONTRIBUTED TO THE REPORTED EVENT. CURRENT CONTROLS FOR MANUFACTURING ARE IN PLACE TO ASSURE FUNCTIONALITY AND DEVICE INTEGRITY PRIOR TO SHIPPING. A REVIEW OF PRODUCTION AND QUALITY DOCUMENTATION DID NOT OBSERVE ANY SPECIFIC ISSUES WITH CURRENT MANUFACTURING OR QUALITY CONTROLS THAT MAY HAVE CONTRIBUTED TO THIS INCIDENT. THE INSTRUCTIONS FOR USE (IFU), PROVIDES THE FOLLOWING INFORMATION TO THE USER RELATED TO THE REPORTED FAILURE MODE: HOW SUPPLIED "UPON REMOVAL FROM THE PACKAGE, INSPECT THE PRODUCT TO ENSURE NO DAMAGE HAS OCCURRED." A DEFINITIVE CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. THE RISK ANALYSIS FOR THIS FAILURE MODE WAS REVIEWED, AND IT WAS DETERMINED THAT NO ADDITIONAL RISK MITIGATING ACTIVITY IS REQUIRED. WE WILL CONTINUE TO MONITOR FOR SIMILAR COMPLAINTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

THERE IS NO NEW PATIENT OR EVENT INFORMATION TO REPORT.

Additional Manufacturer Narrative · 1

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. PMA/510(K) # K180291. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

AS REPORTED, A COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER RUPTURED IN THE PATIENT WHEN INFLATED WITH 1.3ML SALINE. THE DEVICE WAS NOT TESTED PRIOR TO USE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS A RESULT OF THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
647678 COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER LKF CANNULA, MANIPULATOR/INJECTOR, UTERINE LKF COOK INC G17870 13126219 00827002178700

Patients

Seq Age Sex Outcome Treatment
1 32 YR