MRIDIAN LINAC
Report
- Report Number
- 3011233554-2021-00001
- Event Type
- Malfunction
- Date Received
- April 29, 2021
- Date of Event
- March 31, 2021
- Report Date
- April 29, 2021
- Manufacturer
- VIEWRAY, INC.
- Product Code
- IYE
- PMA / PMN Number
- K170751
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 003
Narratives
THIS ISSUE DOES NOT AFFECT SYSTEMS USING THE ENGLISH USER INTERFACE (UI); IT ONLY AFFECTS SITES USING THE (B)(6), GERMAN OR ITALIAN UI. INVESTIGATION IDENTIFIED THE ROOT CAUSE AS FOLLOWS: THE TREATMENT PLANNING AND DELIVERY SYSTEM (TPDS) COMMUNICATES THE GATING ALGORITHM SETTINGS TO THE TREATMENT DELIVERY COMPUTATION UNIT (TDCU) USING AN .XML FILE. THE VALUES FOR "% ROI" AND "CONFIDENCE THRESHOLD" ARE DOCUMENTED AS NUMERICAL VALUES. THESE VALUES ARE THEN CONVERTED TO TEXT IN AN .XML FILE. THE COMMUNICATION MESSAGE IS THEN SENT FROM THE DELIVERY UI TO THE TDCU USING THE EUROPEAN DECIMAL NOTATION WHEN TPDS IS SET TO (B)(6), ITALIAN OR GERMAN LANGUAGES. THE NUMERICAL VALUE DOES NOT GET CONVERTED TO THE VALUES EXPECTED BY THE TDCU. THE TDCU IS EXPECTING A NEUTRAL DECIMAL NOTATION AS IN THE ENGLISH UI.
A USER OF VIEWRAY'S MRIDIAN LINAC SYSTEM REPORTED AN OCCURRENCE WHERE A CHANGE TO THE REGION OF INTEREST (ROI) MADE IN THE "EDIT TRACKING" WINDOW WAS NOT REFLECTED IN TREATMENT DELIVERY IN THE (B)(6) LANGUAGE USER'S INTERFACE (UI) VERSION. THE ROI ALWAYS DEFAULTS TO 5%. THE SITE FURTHER REPORTED THE CONFIDENCE THRESHOLD DOES NOT WORK AT ANYTHING OTHER THAN DEFAULT 50% IN (B)(6) LANGUAGE UI VERSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 645148 | MRIDIAN LINAC | RADIATION THERAPY | IYE | VIEWRAY, INC. | 20000-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |