FDA Adverse Event Malfunction Summary report: N

HUDSON VENTILATOR CIRCUIT,PEDIATRIC W/WATER TR

MDR report key: 11747700 · Received April 29, 2021

Report

Report Number
3004365956-2021-00167
Event Type
Malfunction
Date Received
April 29, 2021
Date of Event
February 15, 2021
Report Date
April 14, 2021
Product Code
CAI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QN#(B)(4). COMPLAINT VERIFICATION TESTING COULD NOT BE PERFORMED AS NO SAMPLE WAS RETURNED FOR ANALYSIS. A VISUAL INSPECTION OF THE PRODUCT INVOLVED WAS PERFORMED ON A PICTURE PROVIDED BY THE CUSTOMER. IT CAN OBSERVE THE PART NUMBER 10714-01 (4 NYLON TIE STRAP) LOOSE FROM 22 MM ADAPTOR. HOWEVER, THE PICTURE IS NOT CLEAR ENOUGH TO DETERMINE ALLEGED ISSUE "LEAK-OTHER". A DEVICE HISTORY RECORD REVIEW WAS PERFORMED, AND NO RELEVANT FINDINGS WERE IDENTIFIED. WITHOUT THE DEVICE TO EVALUATE THE COMPLAINT COULD NOT BE CONFIRMED AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE. IF THE SAMPLE BECOMES AVAILABLE AT A LATER DATE A FOLLOW UP REPORT WILL BE SUBMITTED WITH INVESTIGATION RESULTS.

Description of Event or Problem · 1

IT WAS REPORTED "PRIOR TO USE, THEY TESTED THE CIRCUIT AND FOUND OUT AN AIR ESCAPE LOCATED IN THE BAND THAT IS ATTACHED TO THE FILTER, IT LOOKS BROKEN". NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641217 HUDSON VENTILATOR CIRCUIT,PEDIATRIC W/WATER TR CIRCUIT, BREATHING (W CONNECTO CAI 74F1900018

Patients

Seq Age Sex Outcome Treatment
1