FDA Adverse Event Malfunction Summary report: N

ALERE AFINION AS100 ANALYZER

MDR report key: 11744655 · Received April 29, 2021

Report

Report Number
3003045237-2021-00002
Event Type
Malfunction
Date Received
April 29, 2021
Date of Event
April 23, 2021
Report Date
May 25, 2023
Manufacturer
ABBOTT DIAGNOTICS TECHNOLOGIES AS
Product Code
JQT
PMA / PMN Number
K180296
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INSTRUMENT WILL BE RETURNED TO THE MANFUCATURER FOR INVESTIGATION. THE INCIDENT RELATES TO THE OLD POWER SUPPLY. THE INSTRUMENT WAS RETURNED BUT UNFORTUNATELY THE POWER SUPPLY WASN'T INCLUDED IN THE RETURN PACKAGE. THERE WAS NO INDICATION OF SOOT OR BURN ON THE INSTRUMENT AND IT STARTED UP WHEN USED WITH A NEW POWER SUPPLY. AFINION CRP CONTROL WAS TESTED AND PASSED, AS WELL. IN CONCLUSION, THE ISSUE IS CONSIDERED RELATED TO THE POWER SUPPLY USED AT THE MOMENT OF THE EVENT.

Description of Event or Problem · 0

CUSTOMER REPORTED THAT THERE HAD BEEN SPARKS COMING FROM THE CONNECTION BETWEEN THE AFINION S100 INSTRUMENT (SERIAL NUMBER (B)(6)) AND THE POWER SUPPLY (SERIAL NUMBER (B)(6)). CUSTOMER REPORTED THAT THEY THOUGHT THE FAULT WAS WITH THE POWER SUPPLY AS THEY HAD TRIED TO POWER ON THE INSTRUMENT USING A POWER SUPPLY FROM ANOTHER INSTRUMENT AND IT SEEMED TO WORK. CUSTOMER HAS NOT USED THE INSTRUMENT SINCE THE INCIDENT. CUSTOMER CONFIRMED NO ONE WAS HARMED BY THE OBSERVED SPARKS AND NO ONE HAD EXPERIENCED STATIC ELECTRIC ISSUES AT ANY TIME. POWER SUPPLY WILL BE RETURNED TOGETHER WITH THE INSTRUMENT.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WILL BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE INCIDENT RELATES TO THE OLD POWER SUPPLY.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THERE HAD BEEN SPARKS COMING FROM THE CONNECTION BETWEEN THE AFINION S100 INSTRUMENT (SERIAL NUMBER (B)(4)) AND THE POWER SUPPLY (SERIAL NUMBER (B)(4)). CUSTOMER REPORTED THAT THEY THOUGHT THE FAULT WAS WITH THE POWER SUPPLY AS THEY HAD TRIED TO POWER ON THE INSTRUMENT USING A POWER SUPPLY FROM ANOTHER INSTRUMENT AND IT SEEMED TO WORK. CUSTOMER HAS NOT USED THE INSTRUMENT SINCE THE INCIDENT. CUSTOMER CONFIRMED NO ONE WAS HARMED BY THE OBSERVED SPARKS AND NO ONE HAD EXPERIENCED STATIC ELECTRIC ISSUES AT ANY TIME. POWER SUPPLY WILL BE RETURNED TOGETHER WITH THE INSTRUMENT.)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
646802 ALERE AFINION AS100 ANALYZER ANALYZER JQT ABBOTT DIAGNOTICS TECHNOLOGIES AS 1116053 10185355

Patients

Seq Age Sex Outcome Treatment
1 Unknown