FDA Adverse Event Malfunction Summary report: N

PEN NDL 32GA 4MM 14BAG 700CASE JP

MDR report key: 11742819 · Received April 28, 2021

Report

Report Number
9616656-2021-00448
Event Type
Malfunction
Date Received
April 28, 2021
Date of Event
April 1, 2021
Report Date
June 4, 2021
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 4/14/2021. H.6. INVESTIGATION: ONE OPEN 32G X 4MM PEN NEEDLE SAMPLE AND TWO PHOTOS WERE RETURNED FROM LOT. NO. 0203701, CAT. NO.320136. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLE AND PHOTOS AND DAMAGE TO THE HUB WAS OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. THIS ISSUE WAS CAUSED BY AN INTERFERENCE FIT BETWEEN THE HUB AND COVER.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PEN NDL 32GA 4MM 14BAG 700CASE JP DID NOT ATTACH DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED THAT THE PATIENT COULD NOT ATTACH THE PEN NEEDLE TO THE PEN (POSSIBLY DUE TO A SCREW THREAD-RELATED DEFECT)."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PEN NDL 32GA 4MM 14BAG 700CASE JP DID NOT ATTACH DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED THAT THE PATIENT COULD NOT ATTACH THE PEN NEEDLE TO THE PEN (POSSIBLY DUE TO A SCREW THREAD-RELATED DEFECT)."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632643 PEN NDL 32GA 4MM 14BAG 700CASE JP HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 0203701

Patients

Seq Age Sex Outcome Treatment
1