PEN NDL 32GA 4MM 14BAG 700CASE JP
Report
- Report Number
- 9616656-2021-00448
- Event Type
- Malfunction
- Date Received
- April 28, 2021
- Date of Event
- April 1, 2021
- Report Date
- June 4, 2021
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 4/14/2021. H.6. INVESTIGATION: ONE OPEN 32G X 4MM PEN NEEDLE SAMPLE AND TWO PHOTOS WERE RETURNED FROM LOT. NO. 0203701, CAT. NO.320136. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLE AND PHOTOS AND DAMAGE TO THE HUB WAS OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. THIS ISSUE WAS CAUSED BY AN INTERFERENCE FIT BETWEEN THE HUB AND COVER.
IT WAS REPORTED THAT A PEN NDL 32GA 4MM 14BAG 700CASE JP DID NOT ATTACH DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED THAT THE PATIENT COULD NOT ATTACH THE PEN NEEDLE TO THE PEN (POSSIBLY DUE TO A SCREW THREAD-RELATED DEFECT)."
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT A PEN NDL 32GA 4MM 14BAG 700CASE JP DID NOT ATTACH DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED THAT THE PATIENT COULD NOT ATTACH THE PEN NEEDLE TO THE PEN (POSSIBLY DUE TO A SCREW THREAD-RELATED DEFECT)."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632643 | PEN NDL 32GA 4MM 14BAG 700CASE JP | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 0203701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |