COLLEAGUE CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-05567
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- March 1, 2007
- Report Date
- March 2, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON APR 12, 2007. THE FACILITY REPORTED A PUMP WITH FAILURE CODE 810:04. IT IS UNKNOWN WHEN THIS OCCURRED. EVALUATION SUMMARY:DURING PRODUCT EVALUATION, THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) VALUES WERE FOUND TO BE OUT OF SPECIFICATION. FAILURE CODE 810:04 WAS MANIFESTED AS A RESULT OF THIS CONDITION. THE AIL PCB WAS RECALIBRATED.
DURING PRODUCT EVALUATION, THE PUMP'S AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) WAS FOUND TO BE OUT OF SPECIFICATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.