FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1173648 · Received September 25, 2008

Report

Report Number
6000001-2007-05567
Event Type
Malfunction
Date Received
September 25, 2008
Date of Event
March 1, 2007
Report Date
March 2, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON APR 12, 2007. THE FACILITY REPORTED A PUMP WITH FAILURE CODE 810:04. IT IS UNKNOWN WHEN THIS OCCURRED. EVALUATION SUMMARY:DURING PRODUCT EVALUATION, THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) VALUES WERE FOUND TO BE OUT OF SPECIFICATION. FAILURE CODE 810:04 WAS MANIFESTED AS A RESULT OF THIS CONDITION. THE AIL PCB WAS RECALIBRATED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION, THE PUMP'S AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) WAS FOUND TO BE OUT OF SPECIFICATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment