FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1173646 · Received September 25, 2008

Report

Report Number
6000001-2007-05449
Event Type
Malfunction
Date Received
September 25, 2008
Date of Event
March 1, 2007
Report Date
March 5, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
2005-047-MD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON APR 04 2007. EVALUATION SUMMARY:DURING PRODUCT EVALUATION, A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS OBSERVED. FAILURE CODE 402:317:858:0000 IN THE EVENT HISTORY CONFIRMS THE DEFECTIVE UIM PCB. THIS FAILURE CODE IS MANIFESTED AS A RESULT OF THE UIM PCB BEING DEFECTIVE. THE UIM PCB WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA. THE FACILITY REPORTED THE INFUSION PUMP WITH A FAILURE CODE 402:317:858:0000.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO BAXTER FOR SERVICE. DURING TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH A DEFECTIVE USER INTERFACE MECHANISM PRINTED CIRCUIT BOARD (UIM PCB). ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAS BEEN REPORTED RELATED TO THIS DEVICE. NO ADDITIONAL INFORMATION IS KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1