GENESIS II SPC F/E SZ GDE
Report
- Report Number
- 1020279-2021-03566
- Event Type
- Malfunction
- Date Received
- April 27, 2021
- Date of Event
- April 7, 2021
- Report Date
- July 7, 2021
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- MBH
- PMA / PMN Number
- K121393
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PHYSICIAN
Narratives
H3, H6: THE ASSOCIATED DEVICE, USED IN TREATMENT, WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THE STATED FAILURE MODE. THE DEVICE GUIDE MECHANISM IS BROKEN RENDERING DEVICE INOPERABLE. THE DEVICE WAS MANUFACTURED IN 2013 AND SHOWS SIGNS OF EXTENSIVE USE. A COMPLAINT HISTORY REVIEW FOUND RELATED FAILURES; THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS AND ASSESSED FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.
IT WAS REPORTED THAT DURING SURGERY OF A TKA A GENESIS II SPC F/E SZ GDE, BROKE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 626421 | GENESIS II SPC F/E SZ GDE | PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY | MBH | SMITH & NEPHEW, INC. | 17KCT0001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |