FDA Adverse Event Malfunction Summary report: N

R3 LINER IMPACTOR HEAD 38-42MM

MDR report key: 11735797 · Received April 27, 2021

Report

Report Number
1020279-2021-03551
Event Type
Malfunction
Date Received
April 27, 2021
Date of Event
April 7, 2021
Report Date
July 9, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
JDH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H3, H6: THE DEVICE, INTENDED FOR USE IN TREATMENT, WAS NOT RETURNED FOR EVALUATION AND THE REPORTED EVENT COULD NOT BE CONFIRMED. A REVIEW OF COMPLAINT HISTORY DID NOT REVEAL ADDITIONAL COMPLAINTS FOR THE LISTED BATCH. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. WITHOUT THE RETURN OF THE ACTUAL PRODUCT INVOLVED, OUR INVESTIGATION COULD NOT PROCEED. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A R3 LINER IMPACTOR HEAD 38-42MM, BROKE WHEN COMING OFF THE HANDLE. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
626377 R3 LINER IMPACTOR HEAD 38-42MM PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL JDH SMITH & NEPHEW, INC. 07MAP5770

Patients

Seq Age Sex Outcome Treatment
1