FDA Adverse Event Summary report: N

ANTHROS

MDR report key: 117348 · Received August 26, 1997

Report

Report Number
1932356-1997-00004
Date Received
August 26, 1997
Date of Event
August 7, 1997
Manufacturer
THE ANTHROS MEDICAL GROUP
Product Code
INN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: PRODUCT EXTRA WIDE SHOWER CHAIR. THE CHAIRS RIGHT FRONT LOWER LEG FITTING CRACKED ON THE BACK SIDE. THE SAFETY RING DID PERFORM AS DESIGNED. THE CHAIR SHOWS SOME ABUSE ON THE FRONT OF THE UNIT AS IF BEING PUSHED INTO A WALL OR LEDGE. THE CRACK COULD ALSO HAVE BEEN CAUSED BY PUSHING THE CAHIR OVER THE THRESHOLD OF A DOOR. THIS PUTS EXTRA STRESS ON THE FITTINGS. THE CHAIR IS DESIGNED TO BE USED ON LEVEL SURFACES ONLY.

Description of Event or Problem · 1

NA/R WAS PULLING SHOWER CHAIR WITH RESIDENT SITTING ON IT-OUT SHOWER ROOM INTO ALCOVE AND HALLWAY. SHE WAS TURNING SHOWER CHAIR AROUND TO GO FORWARD, SHE HEARD A CREAKING NOISE AND THE RIGHT FRONT WHEEL POPPED OFF THE CHAIR. RESIDENT FELL FORWARD, HITTING HEAD AND RIGHT SHOULDER ON THE FLOOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANTHROS SHOWER CHAIR INN THE ANTHROS MEDICAL GROUP * *

Patients

Seq Age Sex Outcome Treatment
1 70 YR