COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-03300
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- January 24, 2007
- Report Date
- January 24, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 29 2007. EVALUATION SUMMARY:EVALUATION WAS PERFORMED AND THE CONDITION OF FAILURE CODE 500:320:844:0000 WAS CONFIRMED. FAILURE CODE 500 OCCURS WHEN THE KEY BECOMES STUCK OR IS PRESSED MORE THAN 50 SECONDS AFTER THE PUMP HAS BEEN POWERED ON. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
DURING SERVICE BY BAXTER, FAILURE CODE 500:320:844:0000 WAS FOUND IN THE EVENT HISTORY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.