FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
MDR report key: 1173324
·
Received September 25, 2008
Report
- Report Number
- 6000001-2007-04081
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- January 8, 2007
- Report Date
- January 8, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION WAS PERFORMED AND THE CONDITION OF FAILURE CODE 570 WAS CONFIRMED IN THE EVENT HISTORY. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES CAUSED THE FAILURE CODE 570. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
Description of Event or Problem · 1
DURING SERVICE BY BAXTER, FAILURE CODE 570 WAS FOUND IN THE EVENT HISTORY OF THE DEVICE. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |