FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9600
MDR report key: 1173204
·
Received September 19, 2008
Report
- Report Number
- 1720753-2008-26020
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- September 3, 2008
- Report Date
- September 18, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS, INC
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYS AND REPLACED THE BRAKE AND PAD SPRING. SYS OPERATES AS INTENDED.
Description of Event or Problem · 1
CUSTOMER REPORTED THE C-ARM ROTATION IS NOT LOCKING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS, INC | 9600 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|