FDA Adverse Event Malfunction Summary report: N

ROTATING IV POLE

MDR report key: 11731527 · Received April 27, 2021

Report

Report Number
11731527
Event Type
Malfunction
Date Received
April 27, 2021
Date of Event
April 14, 2021
Report Date
April 20, 2021
Manufacturer
OHMEDA MEDICAL
Product Code
FOX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
UT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

NURSE REPORTED SHE WAS PUTTING AN OXYGEN FLOW MONITOR ON THE WALL BEHIND AN INFANT WARMER IN THE WELL-BABY NURSERY. SHE MOVED THE OHMEDA MEDICAL ROTATING IV POLE OUT OF THE WAY (WHICH IS ATTACHED TO THE INFANT WARMER) AND IT ALL SNAPPED OFF AND FELL. NURSE CHECKED THE SCREWS THAT SECURE THE ROTATING IV POLE TO THE WARMER AND THE METAL HOLDING IT TOGETHER HAD BROKEN. THE WEIGHT LIMIT FOR THE POLE IS 9 KG. THE TWO INFUSION DEVICES ON THE ROTATING IV POLE HAD A COMBINED WEIGHT OF 4.21 KG WHICH IS LESS THAN THE WEIGHT LIMIT. IF THERE HAD BEEN A BABY IN THAT WARMER IT COULD HAVE SEVERELY INJURED THE BABY. CLINICAL ENGINEERING COMPLETED AN INVESTIGATION OF THIS EVENT AND FOUND THAT THE INTERNAL SLEEVE IN WHICH THE BOLT THAT SECURES CLAMPING PART TO THE POLE WAS SHEARED OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630135 ROTATING IV POLE STAND, INFUSION FOX OHMEDA MEDICAL LAU14256889

Patients

Seq Age Sex Outcome Treatment
1