FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1173025 · Received September 19, 2008

Report

Report Number
1720753-2008-25996
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 28, 2008
Report Date
September 18, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE PS3 POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VERTICAL LIFT COLUMN ON THE 9800 SYSTEM WOULD NOT WORK. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS NA NA

Patients

Seq Age Sex Outcome Treatment
1