Malfunction
Summary report: N
GE OEC 9900
MDR report key: 1173024
·
Received September 19, 2008
Report
- Report Number
- 1720753-2008-25964
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- August 21, 2008
- Report Date
- September 17, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE FILAMENT DRIVER BOARD, POWER MOTOR BOARD, AND THE FLUORO FUNCTIONS BOARD. SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
CUSTOMER REPORTED AN INTERMITTENT FAST STOOP ACTIVATED ERROR MESSAGE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | 9900 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |